RIA16-BA2A
Field Indicator, 5 Digit, 26 mm, Current, -40 to 80 °C, RIA16 Series
⚠️ Reference pricing provided. In case of supply shortages, we will connect you with our trusted procurement partners to ensure your project's continuity.
- Manufacturer: ENDRESS+HAUSER
- Product type: Digital Panel Meters
- SVHC: To Be Advised
- Meter Range: 4mA to 20mA
- Digit Height: 26mm
- Product Range: RIA16 Series
- Meter Function: Current
- Panel Cutout Width: -
- Supply Voltage Max: -
- Supply Voltage Min: -
- Panel Cutout Height: -
- No. of Digits / Alpha: 5
- Operating Temperature Max: 80°C
- Operating Temperature Min: -40°C
| Delivery and price | |
|---|---|
| Units per pack | 1 |
| Price | 377.0 € |
| Current stock | 10+ |
| Lead time | 30 days |
TI00144R/09/EN/06.23-00 71613277 2023-02-13
Products
Solutions
Services
## Technical Information RIA16
## Loop-powered field indicator
## Application
- Oil & gas
- Petrochemical industry
- System and apparatus engineering
- Outdoor applications
- Laboratory facilities
- Process data acquisition and monitoring
- Optional: aluminum housing for Ex application
## Your benefits
- Loop-powered indicator
- 5-digit LC display, character height 26 mm (1.02 in)
- Trend bar graph in increments of 10%
- Measuring range display from -19999 to 99999
- Digital limit switch
- Freely programmable units
- 3-key operation
- Approvals: ATEX, FM, CSA, IECEx, NEPSI, UK CA, DNV GL
- Parameter configuration via interface using PC Software
- Parameter configuration without power connection using setup box
RIA16
## Function and system design
## Measuring principle
**==> picture [332 x 143] intentionally omitted <==**
**----- Start of picture text -----**<br>
Hazardous area Non-hazardous<br>Zone 1, 2 EX EX area<br>Transmitter<br>Power supply<br>4...20 mA<br>**----- End of picture text -----**<br>
**==> picture [22 x 4] intentionally omitted <==**
**----- Start of picture text -----**<br>
A0052354<br>**----- End of picture text -----**<br>
_ 1 Example of an application of the field indicator_
The indicator records an analog measuring signal and shows this on the display. The LC display shows the current measured value in digital form and as a bar graph indicating a limit value violation. The indicator is looped into the 4 to 20 mA circuit and gets the required energy from there.
Measuring system Microcontroller-controlled indicator in single-chamber field housing with illuminated LC display. The measuring range, decimal point and offset of the indicator can be configured comfortably by means of three keys in the device with the housing open or by means of a PC with the FieldCare PC software.
## Input
Measured variable Current Measuring range 4 to 20 mA (reverse polarity protection) Input signal • Line voltage drop < 4 V at 3 to 22 mA • Maximum line voltage drop < 6 V at maximum short-circuit current 200 mA
## Output
||Output|
|---|---|
|Output signal|Switching output|
||Digital limit switch|
||Passive, open collector:|
||• Imax= 200 mA|
||• Umax= 35 V|
||• Ulow/max≤ 2 V at 200 mA|
||• Maximum reaction time to limit value = 250 ms|
||• Temperature range: –20 to +80 °C (–4 to +176 °F)|
|Signal on alarm|• No measured value visible on the LC display, no background illumination.|
||• Open Collector inactive.|
|Transmission behavior|The indicator allows the HART® transmission protocol to pass unimpeded.|
Endress+Hauser
2
RIA16
## Power supply
## Terminal assignment
**==> picture [326 x 287] intentionally omitted <==**
**----- Start of picture text -----**<br>
active<br>+ [-]<br>2 3 -<br>OC<br>Terminal 2 and 3<br>Open Collector +-1a1b<br>3 2<br>passive<br>+ [-][ 1a1b] + [- 1a1b]<br>- loop power - loop power<br>+ supply + supply<br>+<br>I<br>Y<br>- +<br>+ - + -<br>I I<br>Y Y<br>**----- End of picture text -----**<br>
**==> picture [22 x 4] intentionally omitted <==**
**----- Start of picture text -----**<br>
A0051980<br>**----- End of picture text -----**<br>
_ 2 Field indicator terminal assignment_
|Terminal|Terminal assignment|Input and output|
|---|---|---|
|+|Measuring signal (+) 4 to 20 mA|Signal input|
|-|Measuring signal (-) 4 to 20 mA|Signal input|
|1a, 1b|Terminal for further instrumentation|Support terminal|
|2|Digital limit switch (collector)|Switch output|
|3|Digital limit switch (emitter)|Switch output|
## Supply voltage
## Cable entries
Power is supplied via the 4 to 20 mA current loop.
The following cable entries are available:
- 2x thread NPT1/2
- 2x thread M16
The plastic housing includes up to 5 cable entries. In the housing as supplied to the customer, 3 of these remain closed. If necessary, you can open these using a suitable tool.
The aluminum housing also includes 5 cable entries, 3 of which are closed with blind plugs.
Endress+Hauser
3
RIA16
## Performance characteristics
||Performance characteristics|
|---|---|
|Reference operating|T = 25 °C (77 °F)|
|conditions||
|Maximum measured error|< 0.1% of scaled display range|
|Influence of ambient|Effect on the accuracy when ambient temperature changes by 1 K (1.8 °F): 0.01%|
|temperature (temperature||
|drift)||
## Mounting
||Mounting|
|---|---|
|Mounting location|Wall or pipe mounting (see ’Accessories’ )|
|Orientation|No restrictions: The device orientation is determined by the readability of the display.|
|Operating altitude|Up to 2 000 m (6 561.7 ft) above sea level|
||Environment|
|Ambient temperature range|–40 to +80 °C (–40 to +176 °F)|
||–20 to +80 °C (–4 to +176 °F) when the Open Collector output is used|
||The display can react slowly for temperatures < –20 °C (–4 °F).<br>Readability of the display cannot be guaranteed at temperatures < –30 °C (–22 °F).|
|Storage temperature|–40 to +80 °C (–40 to +176 °F)|
|Electrical safety|As per IEC 61010-1,|
||UL61010-1,|
||CSA C22.2 No. 1010.1-92|
|Climate class|As per IEC 60654-1, Class C|
|Degree of protection|IP 67, NEMA 4X (not UL rated)|
|Shock resistance|3g/2 to 150 Hz as per IEC 60068-2-6|
|Condensation|Permitted|
|Installation category|1 according to IEC 61010|
|Pollution degree|2|
|Overvoltage category|II|
|Electromagnetic|CE conformity|
|compatibility (EMC)|Electromagnetic compatibility in accordance with all the relevant requirements of the IEC/EN 61326|
||series and NAMUR Recommendation EMC (NE21). For details, refer to the Declaration of|
||Conformity.|
Endress+Hauser
4
RIA16
Maximum measured error <1% of measuring range.
Interference immunity as per IEC/EN 61326 series, industrial requirements Interference emission as per IEC/EN 61326 series, Class B equipment
Connection of the functional grounding may be needed for functional purposes. Compliance with the electrical codes of individual countries is mandatory.
## Mechanical construction
## Design, dimensions
**==> picture [387 x 165] intentionally omitted <==**
**----- Start of picture text -----**<br>
96 (3.78)<br>A-A<br>A<br>1<br>A<br>199 (7.83) 163 (6.42)<br> A0011162<br> 3 Dimensions in mm (in)<br>1 Bore hole for mounting directly to wall or optional mounting plate with 4 screws ⌀ 5 mm (0.2 in)<br>133 (5.24) 115 (4.53)<br>158 (6.22)<br>**----- End of picture text -----**<br>
|Weight|• Plastic housing: approximately 500 g (1.1 lb)<br>• Aluminum housing: approximately 1.7 kg (3.75 lb)|• Plastic housing: approximately 500 g (1.1 lb)<br>• Aluminum housing: approximately 1.7 kg (3.75 lb)|
|---|---|---|
|Materials|Housing|Nameplate|
||Fiber-glass reinforced plastic PBT-GF30|Laser marking|
||Optional: Aluminum (AlSi12, AC-44100 or AlSi10Mg(Fe), AC-43400)|Laser-writable foil, polyester|
||||
|Terminals|Cables / wires up to 2.5 mm2(14 AWG) plus ferrule<br>Human interface||
|Operating concept|3-key operation (-/+/E) integrated in device; access when housing is open||
Endress+Hauser
5
RIA16
On-site operation
**==> picture [203 x 96] intentionally omitted <==**
**----- Start of picture text -----**<br>
1 1b<br>1a<br>5<br>2<br>4<br>3<br>6<br>**----- End of picture text -----**<br>
**==> picture [22 x 4] intentionally omitted <==**
**----- Start of picture text -----**<br>
A0011163<br>**----- End of picture text -----**<br>
- _ 4 LC display of the field indicator_
- _1 Bar graph display in increments of 10% with indicators for measurement range undershoot (item 1a) and overshoot (item 1b)_
- _2 Measured value display, digit height 26 mm (1.02 in)_
- _3 14-segment display for units and messages_
- _4 "Programming disabled" symbol_
- _5 Unit "%"_
- _6 "Failure" warning symbol_
- Display range
- -19999 to 99999
- Offset
- -19999 to 99999
- Signalization Measuring range overshoot/undershoot
- Limit value violation Limit value overshoot/undershoot
## Remote operation
## Parameter configuration
The device can be configured with the FieldCare PC software. FieldCare Device Setup is included in the Commubox FXA291 and TXU10-AC scope of delivery (see ‘Accessories’) or can be downloaded free of charge at www.endress.com.
## Interface
Configuration interface on the device; connection to the PC via interface cable (see 'Accessories').
## Configurable device parameters (selection)
Measuring dimension, measuring ranges (linear/squared), setup lock with user code, failure mode, digital filter (damping), offset, limit value (min/max/alarm), user-configurable alarm limit values
Endress+Hauser
6
RIA16
**==> picture [331 x 183] intentionally omitted <==**
**----- Start of picture text -----**<br>
1 2 3<br>USB ele) a ©@O<br>— — Yi XK<br>{| ) |) —&<br>> \ Ya Qj<br>SQW _®_ GEE (665<br> A0051930<br>**----- End of picture text -----**<br>
- _ 5 Configuring the field indicator using an interface adapter_
- _1 PC configuration software_
- _2 Configuration kit USB box_
- _3 Field indicator_
## Certificates and approvals
Current certificates and approvals for the product are available at www.endress.com on the relevant product page:
1. Select the product using the filters and search field.
2. Open the product page.
3. Select Downloads.
## UL approval
More information under UL Product iq™, search for keyword "E225237")
## Ordering information
Detailed ordering information is available from your nearest sales organization www.addresses.endress.com or in the Product Configurator at www.endress.com:
1. Select the product using the filters and search field.
2. Open the product page.
3. Select Configuration.
Product Configurator - the tool for individual product configuration
- Up-to-the-minute configuration data
- Depending on the device: Direct input of measuring point-specific information such as measuring range or operating language
- Automatic verification of exclusion criteria
- Automatic creation of the order code and its breakdown in PDF or Excel output format
- Ability to order directly in the Endress+Hauser Online Shop
## Accessories
Accessories currently available for the product can be selected via the Product Configurator at www.endress.com:
1. Select the product using the filters and search field.
2. Open the product page.
Endress+Hauser
7
RIA16
3. Select Spare parts & Accessories.
|Device-specific accessories|Designation||Description|
|---|---|---|---|
||Wall/pipe mounting|kit|For pipes ⌀1-5" , different variants for plastic and aluminum housings|
||Weather protection c|over|163.9 (6.45)<br>274 (10.79)<br>272(10.71)<br>140°<br>30<br>(1.18)<br>95°<br>180 (7.09)<br>A0021548<br>_ 6_<br>_Dimensions in mm (in)_|
|||||
|Communication-specific<br>accessories|Designation|||
||Interface cable|Commubox TXU10 incl. FieldCare Device Setup and DTM Library||
|||Commubox FXA291 incl. FieldCare Device Setup and DTM Library||
## Supplementary documentation
The following types of documentation are available on the product pages and in the Download Area of the Endress+Hauser website (www.endress.com/downloads) (depending on the selected device version):
|Document<br>Pur|pose and content of the document|
|---|---|
|Technical Information (TI)<br>Plan<br>The<br>an o<br>the|ning aid for your device<br>document contains all the technical data on the device and provides<br>verview of the accessories and other products that can be ordered for<br>device.|
|Brief Operating Instructions (KA)<br>Guid<br>The<br>from|e that takes you quickly to the 1st measured value<br>Brief Operating Instructions contain all the essential information<br>incoming acceptance to initial commissioning.|
|Operating Instructions (BA)<br>You<br>The<br>vari<br>inco<br>and<br>disp|r reference document<br>Operating Instructions contain all the information that is required in<br>ous phases of the life cycle of the device: from product identification,<br>ming acceptance and storage, to mounting, connection, operation<br>commissioning through to troubleshooting, maintenance and<br>osal.|
|Description of Device Parameters<br>(GP)<br>Refe<br>The<br>para<br>over|rence for your parameters<br>document provides a detailed explanation of each individual<br>meter. The description is aimed at those who work with the device<br>the entire life cycle and perform specific configurations.|
|Safety Instructions (XA)<br>Dep<br>devi<br>Instr<br>|ending on the approval, Safety Instructions (XA) are supplied with the<br>ce. The Safety Instructions are an integral part of the Operating<br>uctions.<br>Information on the Safety Instructions (XA) that are relevant for<br>the device is provided on the nameplate.|
|Supplementary device-dependent<br>documentation (SD/FY)<br>Alw<br>supp<br>inte|ays comply strictly with the instructions in the relevant<br>lementary documentation. The supplementary documentation is an<br>gral part of the device documentation.|
Endress+Hauser
8
*71613277* 71613277
www.addresses.endress.com
Updated at June 9, 2026
About Novapart
Novapart is a B2B electronic component broker specialising in stock shortages and cost reduction. We source hard-to-find parts and identify compliant alternatives across a catalogue of 410,000+ components from 500+ manufacturers.
Learn more →Stock Shortage Specialist
When a component is unavailable, discontinued or has an unacceptable lead time, we tap into our network of vetted European and Asian distributors to source what you need — without compromising on quality or traceability.
Request a quote →Compliant Alternatives
We identify pin-to-pin, electrically equivalent substitutes that meet the same certifications (RoHS, AEC-Q100, REACH) as your original specification — validated against datasheets, not just part numbers. Often at a lower cost.
BOM Analysis service →