MDS-100BPS12 BA
POWER SUPPLY, AC-DC, 12V, 6.6A
⚠️ Reference pricing provided. In case of supply shortages, we will connect you with our trusted procurement partners to ensure your project's continuity.
- Manufacturer: DELTA ELECTRONICS / POWER
- Product type: AC / DC Open Frame Power Supplies
- Product Range: MDS Series
- No. of Outputs: 1 Output
- Input Voltage VAC: 90V AC to 264V AC
- Power Supply Approvals: ITE & Medical
- Power Supply Output Type: Fixed
- Output Current - Output 1: 6.6A
- Output Voltage - Output 1: 12V
- Power Rating (Forced Cooling): 100W @ 10CFM
- Power Rating (Convection Cooling): 80W
| Delivery and price | |
|---|---|
| Units per pack | 100 |
| Price | 26.96 € |
| Current stock | 25+ |
| Lead time | 7 days |
**TECHNICAL DATASHEET**
## **MDS Medical AC-DC Open Frame** 12Volt, 100Watt / MDS-100BPS12 B
## **100BPS12 B**
## **Highlights & Features**
- Safety Approvals to IEC 60601-1 3.1rd ed. & IEC 60950-1
- Compliant with IEC 60601-1-2 4th Ed. Requirements
- Compact size 2”x4”
- Low touch current (<0.1mA Normal & <0.3mA single fault)
- Over-Voltage/Load/Temperature & Short Circuit protections
- 1 Million Hours MTBF
- 2 x MOPP (means of patient protection)
- 3 years warranty
## **Safety Standards**
CB Certified for worldwide use
**Model Number:** MDS-100BPS12 B **Unit Weight:** 150 grams (5.3 Ounces) **Dimensions (W x L x H):** 50.8x101.6x31.8 mm 2.0x4.0x1.25 in
## **General Description**
The MDS series of embedded power supply come with universal AC input at 90Vac to 264Vac. Other features include low touch current, risk management report available and the electric shock protection comply with 2 x MOPP. The MDS series is certified for EMC standards according to EN 55011 for industrial, scientific and medical (ISM) radio-frequency equipment and EN 55022 for Information Technology Equipment (ITE) radio-frequency equipment. In addition, only recognized Japanese capacitors are used.
The MDS series of embedded power supply come with both medical and ITE safety approvals including UL/cUL/CQC /CE and CB certification and are fully compliant with RoHS Directive 2011/65/EU for environmental protection.
## **Model Information**
## Medical AC-DC Open Frame
**==> picture [522 x 134] intentionally omitted <==**
**----- Start of picture text -----**<br>
Model Number Input Voltage Output Voltage Conversion Current Output Forced Air Current Output<br>MDS-100BPS12 B 90-264Vac 12Vdc 6.6A 8.4A<br>Model Numbering<br>MDS - 100 BPS 12 B<br>Delta Medical Max wattage in the product series. Family Code Output Voltage<br>power Supply Maybe lower at some voltage. Single Output:<br>100 → 100W 12 for 12V<br>**----- End of picture text -----**<br>
## **Model Numbering**
All parameters are specified at 25°C ambient unless otherwise noted. 1 **www.DeltaPSU.com (March 2019, Rev. 06)**
**TECHNICAL DATASHEET**
**MDS Medical AC-DC Open Frame** 12Volt, 100Watt / MDS-100BPS12 B
## **Specifications**
Input Ratings / Characteristics
|**Specifications**<br>Input Ratings / Characteristics||
|---|---|
|Nominal Input Voltage|100-240Vac|
|Input Voltage Range|90-264Vac|
|Nominal Input Frequency|50-60Hz|
|Input FrequencyRange|47-63Hz|
|Input Current (max)|2.5A @ 100Vac, 1.25A @ 240Vac|
|Efficiency (typ.)|87%, Reference Fig.1|
|StandbyPower(max)|0.3W|
|Inrush Current (typ.)|40A @ 115Vac, 80A @ 230Vac|
|Touch Current(max)|0.1mA@264Vac NC1), 0.3mA@264Vac SFC2)|
|Earth Leakage Current(max)|0.15mA@264Vac NC1), 0.3mA@264Vac SFC2)|
1) NC: normal condition
2) SFC: single fault condition
**==> picture [392 x 199] intentionally omitted <==**
**----- Start of picture text -----**<br>
100<br>95<br>Fig.1 Efficiency versus output load<br>90<br>85<br>80<br>75 115Vac Efficiency<br>70 230Vac Efficiency<br>65<br>60<br>55<br>50<br>25% 50% 75% 100%<br>Output load<br>E fficiency (% )<br>**----- End of picture text -----**<br>
## Output Ratings / Characteristics
|Output Ratings / Characteristics||
|---|---|
|Nominal Output Voltage|12Vdc|
|Total Regulation|± 1.5%|
|Output Current|8.33A with 10CFM force air, 6.67A for convection|
|Output Power|100W with 10CFM force air, 80W for convection|
|Line Regulation(max)|±0.5%|
|Load Regulation (max)|±1%|
|Ripple & Noise(typ.)|76mVpk-pk@Full load, Reference Fig. 2|
|Start-up Time(max)|3000ms@115Vac|
|Hold-up Time(min)|10ms@115Vac|
|Dynamic Response(Overshoot & Undershoot O/P Voltage)|± 3%@50-100% load|
*Periodic and Random Deviation
All parameters are specified at 25°C ambient unless otherwise noted. **www.DeltaPSU.com (March 2019, Rev. 06)**
2
**TECHNICAL DATASHEET**
## **MDS Medical AC-DC Open Frame** 12Volt, 100Watt / MDS-100BPS12 B
(a) 115V (b) 230V
Fig. 2 Ripple & Noise example, 20MHz BW
Ripple & Noise measurement circuit
||PSU<br>5cm<br>S~~S~~|PSU<br>5cm<br>S~~S~~|PSU<br>5cm<br>S~~S~~|PSU<br>5cm<br>S~~S~~|||||0.47uF Ceramic<br>10uF Electrolytic|0.47uF Ceramic<br>10uF Electrolytic|0.47uF Ceramic<br>10uF Electrolytic|0.47uF Ceramic<br>10uF Electrolytic|0.47uF Ceramic<br>10uF Electrolytic|0.47uF Ceramic<br>10uF Electrolytic|0.47uF Ceramic<br>10uF Electrolytic||Load<br>||Load<br>||Scope|Scope|Scope|Scope|Scope|Scope|Scope|||||®@|®@|®@|
|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|
||||||||||||||||||||||||||||||||||
||||||||||||||||||||||||||||||||||
||||||||||||||||||||||||||||||||||
||||||||||||||||||||||||||||||||||
||||||Test Fixture||||||||||||||||||||||||||||
|Mechanical|||||||||||||||||||||||||||||||||
||Dimensions(W x Lx H||W x Lx H)||||||||||||||||50.8 x 101.6 x 31.8 mm||50.8 x 101.6 x 31.8 mm|||50.8 x 101.6 x 31.8 mm||||||50.8 x 101.6 x 31.8 mm|||
||Weight(typ.)||||||||||||||||||150grams(5.3 Ounces||||||5.3 Ounces)||||||||
||Terminal||||||||||||||||Input<br>Output||JST 2P<br>JST 4P||||||||||||||
## Environment
|Environment|||
|---|---|---|
|Surrounding Air Temperature|Operating|-10°C to +70°C|
||Storage|-40°C to +85°C|
|Power De-rating||-10°C to +50°C 100% load<br>50°C to 70°C de-ratepower by2.5% / °C,See Fig. 3|
|OperatingHumidity||10-95% RH(Non-Condensing)|
|Operating Altitude||Up to 5,000 meters (up to 16,400 feet or 106-54k Pa)|
|Non-operatingAltitude||Upto 5,575 meters(upto 18,290 feet or 106-50k Pa)|
|Shock Test(Non-Operating)||50G, 11ms, 3 shocks for each direction|
|Vibration(Non-Operating)||5-500Hz, 2.09Grms, 20 minute for each three axis|
All parameters are specified at 25°C ambient unless otherwise noted. **www.DeltaPSU.com (March 2019, Rev. 06)**
3
**TECHNICAL DATASHEET**
**MDS Medical AC-DC Open Frame** 12Volt, 100Watt / MDS-100BPS12 B
## Protections
|Over Voltage(max)|150%, Latch Mode|
|---|---|
|Over Load / Over Current (max)|170% of rated load current, Hiccup Mode,|
||(Non-Latching,Auto-Recovery)|
|Over Temperature|Hiccup Mode,|
||(Non-Latching, Auto-Recovery)|
|Short Circuit|Hiccup Mode,|
||(Non-Latching,Auto-Recovery)|
|Protection Against Shock|Class I with PE* connection|
*PE: Protective Earth
## Reliability Data
|Safety Standards / Directives|Safety Standards / Directives||
|---|---|---|
|Medical Safety||IEC60601-1 3rdand 3.1rdedition CB report<br>IEC60601-1 edition 3.1rd (2012), EN60601-1 (2006) +<br>A11 + A1 + A12, CAN/CSA-C22.2 NO. 60601-1:14,<br>ANSI/AAMI ES60601-1:2005/(R)2012|
|ITE Safety||IEC60950-1 (Ed.2,2005), GB4943.1-2011, GB9254-<br>2008, GB17625.1-2003|
|CE||MDD Directive 93/42/EEC|
|Material and Parts||RoHS Directive 2011/65/EU Compliant|
|Galvanic Isolation|Input to Output|4000 Vac|
||Input to Ground|1500 Vac|
||Output to Ground|500 Vac|
## Safety Standards / Directives
All parameters are specified at 25°C ambient unless otherwise noted. **www.DeltaPSU.com (March 2019, Rev. 06)**
4
**TECHNICAL DATASHEET**
**MDS Medical AC-DC Open Frame** 12Volt, 100Watt / MDS-100BPS12 B
EMC (Compliant with IEC 60601-1-2 4th Ed. Requirements)
|EMC (Compliant with IEC 60601-1-2 4th Ed. Requirements)||
|---|---|
|EMC / Emissions|EN55011/EN55032, FCC Title 47:Class B|
|Harmonic Current Emissions<br>IEC61000-3-2|Meet Class A limit|
|Immunityto||
|Voltage Flicker<br>IEC61000-3-3||
|Electrostatic Discharge<br>IEC61000-4-2|Level 4 Criteria A1)<br>Air Discharge: 15kV<br>Contact Discharge: 8kV|
|Radiated Field<br>IEC61000-4-3|Criteria A1)<br>80MHz-2700MHz, 10V/m AM modulation<br>385MHz-5785MHz, 28V/m Pulse mode and other<br>modulation|
|Electrical Fast Transient / Burst<br>IEC61000-4-4|Level 3 Criteria A1):2kV|
|Surge<br>IEC61000-4-5|Level 3 Criteria A1)<br>Common Mode3): 2kV<br>Differential Mode4): 1kV|
|Conducted<br>IEC61000-4-6|Level 2 Criteria A1)<br>150kHz-80MHz, 3Vrms,6Vrms at ISM bands and<br>Amateur radio bands|
|Power Frequency Magnetic Fields<br>IEC61000-4-8|Criteria A1)<br>Magnetic field strength 30A/m|
|Voltage Dips<br>IEC61000-4-11|Criteria B2)<br>0% UT, 0.5 cycle (10ms)<br>, 0o/45o/90o/135o/180o/225o/270o/315o/360o<br>Criteria B2)<br>0% UT, 1 cycle (20ms), 0o<br>Criteria A1)<br>70% UT, 25 cycle (500ms), 0o<br>Criteria B2)<br>0% UT,250 cycle (5000ms), 0o|
- 1) Criteria A: Normal performance within the specification limits
- 2) Criteria B: Output out of regulation, or shuts down during test. Automatically restore to normal operation after test.
- 3) Asymmetrical: Common mode (Line to earth)
- 4) Symmetrical: Differential mode (Line to line)
All parameters are specified at 25°C ambient unless otherwise noted. **www.DeltaPSU.com (March 2019, Rev. 06)**
5
**TECHNICAL DATASHEET**
**MDS Medical AC-DC Open Frame** 12Volt, 100Watt / MDS-100BPS12 B
## **Block Diagram**
All parameters are specified at 25°C ambient unless otherwise noted. **www.DeltaPSU.com (March 2019, Rev. 06)**
6
**TECHNICAL DATASHEET**
## **MDS Medical AC-DC Open Frame** 12Volt, 100Watt / MDS-100BPS12 B
## **Dimensions**
**W x L x H:** 50.8 x 101.6 x 31.8 mm
**==> picture [19 x 6] intentionally omitted <==**
**----- Start of picture text -----**<br>
TR1<br>**----- End of picture text -----**<br>
TR1 : KANG YANG part number PCH250 mates with KST part number FDFNYD1-250 or equivalent.
## **Notes**
- Dimensions are in mm
- For Open Frame type: There are 4 normal mounting holes. To ensure EMC performance, all mounting holes shall be connected to metal case.
- Method of protective earth connection:
- a. When system (End products’) protective earth connection is connected to system’s (or end products)’ enclosure/cover (afterward named “system” for short).
1. Two mounting holes which are marked ground symbols in mechanical drawing are needed to link system together by conductive screw
2. The cable can be connected system to TR1.
- b. System (End products’) incoming protective earth connection can be connected to TR1 on power supply directly.
All parameters are specified at 25°C ambient unless otherwise noted. **www.DeltaPSU.com (March 2019, Rev. 06)**
7
**TECHNICAL DATASHEET**
## **MDS Medical AC-DC Open Frame** 12Volt, 100Watt / MDS-100BPS12 B
## **Power De-rating**
**==> picture [299 x 203] intentionally omitted <==**
**----- Start of picture text -----**<br>
Output power<br>100%<br>With 10CFM force air<br>80%<br>Convection<br>50%<br>40%<br>0% Ambient<br>-10°C 50°C 70°C Temperature<br>Fig. 3<br>**----- End of picture text -----**<br>
## **Functions**
## Start-up Time
The time required for the output voltage (Vo) to reach 90% of its set value, after the input AC voltage is applied.
## Rise Time
The time required for the output voltage (Vo) to change from 10% to 90% of its steady state value.
## Hold-up Time
Hold up time is the time when the AC input collapses and output voltage retains regulation for a certain period of time. The time required for the output to reach 95% of its set value, after the input voltage is removed.
## ■ **Graph illustrating the Start-up Time, Rise Time, and Hold-up Time**
All parameters are specified at 25°C ambient unless otherwise noted. **www.DeltaPSU.com (March 2019, Rev. 06)**
8
**TECHNICAL DATASHEET**
**MDS Medical AC-DC Open Frame** 12Volt, 100Watt / MDS-100BPS12 B
## Dynamic Response
The power supply output voltage will remain within ±3% of its steady state value, when subjected to a dynamic load change from 50 to 100% of its rated current.
**==> picture [195 x 91] intentionally omitted <==**
**----- Start of picture text -----**<br>
100%<br>Load<br>50%<br>Load i i i T (sec)<br>50ms 50ms 50ms 50ms<br>**----- End of picture text -----**<br>
## Overload & Over current Protections
The power supply’s Overload (OLP) and Over current (OCP) Protections will be activated when output current is between 110% and 170% of IO (Max load). Upon such an occurrence, VO will start to drop. Once the power supply has reached its maximum power limit, the protection will be activated. and the power supply will go into “Hiccup mode” (Auto-Recovery). The power supply will recover once the fault condition causing the OLP and OCP is removed and IO is back within the specified limit.
## Inrush Current
Inrush current is the input current that occurs when the input voltage is first applied. For AC input voltages, the maximum peak value of inrush current will occur during the first half cycle of the applied AC voltage. This peak value decreases exponentially during subsequent cycles of AC voltage.
Additionally, if the IO is <110% but >100% for a prolong period of time (depending on the load), the Over Temperature Protection (OTP) will be activated due to high temperature on critical components. The power supply will then go into hiccup mode until the fault is removed; and, the input voltage is removed, then reapplied.
## Over Temperature Protection
## Overvoltage Protection
The power supply’s overvoltage circuit will be activated when its internal feedback circuit fails. The output voltage shall not exceed its specifications defined on Page 4 under “Protections”. Power supply will latch off, and require removal/re-application of input AC voltage in order to restart.
As mentioned above, the power supply also has Over Temperature Protection (OTP). This is activated when the overload condition persists for an extended duration and the output current is below the overload trigger point but >100% load. In the event of a higher operating condition at 100% load, the power supply will run into OTP when the surrounding air temperature is higher than the operating temperature. When activated, the output voltage will go into hiccup mode until the input voltage is removed; then, reapplied, and the surrounding air temperature drops to its normal operating temperature.
## Short Circuit Protection
The power supply’s output OLP/OCP function also provides protection against short circuits. When a short circuit is applied, the output current will operate in “Hiccup mode”, as shown in the illustration in the OLP/OCP section on this page. The power supply will return to normal operation after the short circuit is removed.
All parameters are specified at 25°C ambient unless otherwise noted. **www.DeltaPSU.com (March 2019, Rev. 06)**
9
**TECHNICAL DATASHEET** ```
**MDS Medical AC-DC Open Frame** 12Volt, 100Watt / MDS-100BPS12 B
## **Certificate**
.
All Delta Medical Power products conform to the European directive 2011/65/EU. RoHS is the abbreviation for “Restriction of the use of certain hazardous substances.
Delta has been certified as meeting the requirement of ISO 13485: 2003 and EN ISO 13485:2012 for the design and manufacture of switching power supply and adaptor for medical device.
In addition to a UL Total Certification Program (TCP) approved client laboratory for IEC60950 and IEC60065. Delta also has participated UL Client Test Data Program (CDTP) for IEC 60601
## **Attention**
Delta provides all information in the datasheets on an “AS IS” basis and does not offer any kind of warranty through the information for using the product. In the event of any discrepancy between the information in the catalog and datasheets, the datasheets shall prevail (please refer to www.DeltaPSU.com for the latest datasheets information). Delta shall have no liability of indemnification for any claim or action arising from any error for the provided information in the datasheets. Customer shall take its responsibility for evaluation of using the product before placing an order with Delta.
Delta reserves the right to make changes to the information described in the datasheets without notice.
All parameters are specified at 25°C ambient unless otherwise noted. **www.DeltaPSU.com (March 2019, Rev. 07)**
10
Updated at February 9, 2023
About Novapart
Novapart is a B2B electronic component broker specialising in stock shortages and cost reduction. We source hard-to-find parts and identify compliant alternatives across a catalogue of 410,000+ components from 500+ manufacturers.
Learn more →Stock Shortage Specialist
When a component is unavailable, discontinued or has an unacceptable lead time, we tap into our network of vetted European and Asian distributors to source what you need — without compromising on quality or traceability.
Request a quote →Compliant Alternatives
We identify pin-to-pin, electrically equivalent substitutes that meet the same certifications (RoHS, AEC-Q100, REACH) as your original specification — validated against datasheets, not just part numbers. Often at a lower cost.
BOM Analysis service →