BA01697D/06/EN/06.22-00 71552010 2022-06-01 Valid as of version 01.01.zz (Device firmware)
## Operating Instructions Picomag
## IO-Link
Electromagnetic flowmeter
Picomag IO-Link
- Make sure the document is stored in a safe place such that it is always available when working on or with the device.
- To avoid danger to individuals or the facility, read the "Basic safety instructions" section carefully, as well as all other safety instructions in the document that are specific to working procedures.
- The manufacturer reserves the right to modify technical data without prior notice. Your Endress+Hauser Sales Center will supply you with current information and updates to these instructions.
Endress+Hauser
2
Picomag IO-Link
Table of contents
## Table of contents
|1|About this document . . . . . . . . . . . . . . . . 5|About this document . . . . . . . . . . . . . . . . 5|
|---|---|---|
|1.1|Document function. . . . . . . . . . . . . . . . . . . . .|5|
|1.2|Symbols used. . . . . . . . . . . . . . . . . . . . . . . . . . 5||
||1.2.1<br>Safety symbols. . . . . . . . . . . . . . . . . .|5|
||1.2.2<br>Electrical symbols. . . . . . . . . . . . . . . .<br>1.2.3<br>Communication-specific symbols. . . . .<br>1.2.4<br>Symbols for certain types of<br>information. . . . . . . . . . . . . . . . . . . .|5<br>5<br>6|
||1.2.5<br>Symbols in graphics. . . . . . . . . . . . . . . 6||
|1.3|Documentation. . . . . . . . . . . . . . . . . . . . . . . .|6|
|1.4|Registered trademarks. . . . . . . . . . . . . . . . . . . 6||
|2|Basic safety instructions . . . . . . . . . . . .|8|
|2.1<br>2.2|Requirements for the personnel. . . . . . . . . . . .<br>Intended use. . . . . . . . . . . . . . . . . . . . . . . . . .|8<br>8|
|2.3<br>2.4<br>2.5<br>2.6<br>2.7|Workplace safety. . . . . . . . . . . . . . . . . . . . . . . 8<br>Operational safety. . . . . . . . . . . . . . . . . . . . . .<br>9<br>Product safety. . . . . . . . . . . . . . . . . . . . . . . . .<br>9<br>IT security. . . . . . . . . . . . . . . . . . . . . . . . . . . .<br>9<br>Device-specific IT security. . . . . . . . . . . . . . . . . 9<br>2.7.1<br>Access via the SmartBlue app. . . . . . . . 9<br>2.7.2<br>Protecting access via a password. . . . .<br>9<br>2.7.3<br>Access via Bluetooth® wireless||
||technology. . . . . . . . . . . . . . . . . . . .|10|
|3|Incoming acceptance and product||
||identification . . . . . . . . . . . . . . . . . . . . .|11|
|3.1<br>3.2|Incoming acceptance. . . . . . . . . . . . . . . . . . .<br>Product identification. . . . . . . . . . . . . . . . . . .|11<br> 11|
||3.2.1<br>Symbols on measuring device. . . . . .|12|
|4<br>4.1|Storage and transport . . . . . . . . . . . . .<br>Storage conditions. . . . . . . . . . . . . . . . . . . . .|13<br> 13|
|4.2<br>4.3|Transporting the product. . . . . . . . . . . . . . . .<br>Packaging disposal. . . . . . . . . . . . . . . . . . . . .|13<br> 13|
|5|Mounting . . . . . . . . . . . . . . . . . . . . . . . . .|14|
|5.1<br>5.2|Mounting requirements. . . . . . . . . . . . . . . . . <br>5.1.1<br>Mounting position. . . . . . . . . . . . . . . <br>Mounting the measuring device. . . . . . . . . . .|14<br> 14<br>14|
|6|Electrical connection . . . . . . . . . . . . . .|16|
|6.1|Electrical safety. . . . . . . . . . . . . . . . . . . . . . .|16|
|6.2|Connecting requirements. . . . . . . . . . . . . . . .|16|
||6.2.1<br>Requirements for connecting cable. . .|16|
||6.2.2<br>Pin assignment, device plug. . . . . . . .|16|
|6.3|Connecting the measuring device. . . . . . . . . .|19|
|6.4|Post-connection check. . . . . . . . . . . . . . . . . .|19|
|7|Operation options . . . . . . . . . . . . . . . . .|20|
|---|---|---|
|7.1|Access to the operating menu via the||
||SmartBlue app. . . . . . . . . . . . . . . . . . . . . . . .|20|
|8|System integration . . . . . . . . . . . . . . . .|21|
|8.1|Overview of device description files. . . . . . . . .|21|
|8.2|Device master file. . . . . . . . . . . . . . . . . . . . .|21|
|9|Commissioning . . . . . . . . . . . . . . . . . . . .|22|
|9.1|Switching ON the measuring device. . . . . . . .|22|
|9.2|Overview of the operating menu. . . . . . . . . . .|22|
|9.3|Configuring the measuring device. . . . . . . . . .|22|
||9.3.1<br>Identification. . . . . . . . . . . . . . . . . .|22|
||9.3.2<br>Configuring system units. . . . . . . . . .|22|
||9.3.3<br>Setting the installation direction and||
||measurement. . . . . . . . . . . . . . . . . .|23|
||9.3.4<br>Configuring the IO modules. . . . . . . .|24|
||9.3.5<br>Totalizer. . . . . . . . . . . . . . . . . . . . . .|29|
||9.3.6<br>Configuring the display. . . . . . . . . . .|30|
||9.3.7<br>Data management. . . . . . . . . . . . . . .|31|
||9.3.8<br>Security. . . . . . . . . . . . . . . . . . . . . .|31|
|9.4|Diagnostics. . . . . . . . . . . . . . . . . . . . . . . . . .|32|
|9.5|System. . . . . . . . . . . . . . . . . . . . . . . . . . . . .|32|
|10|Operation . . . . . . . . . . . . . . . . . . . . . . . . .|33|
|10.1|Offline quick view of configuration. . . . . . . .|33|
|11|Diagnostics and troubleshooting . . .|35|
|11.1|General troubleshooting. . . . . . . . . . . . . . . . .|35|
|11.2|Diagnostic information on local display. . . . . .|36|
||11.2.1 Diagnostic message. . . . . . . . . . . . . .|36|
|11.3|Overview of diagnostic events. . . . . . . . . . . . .|37|
|11.4|Device information. . . . . . . . . . . . . . . . . . . .|37|
|11.5|Firmware history. . . . . . . . . . . . . . . . . . . . .|38|
|12|Accessories . . . . . . . . . . . . . . . . . . . . . . .|39|
|13|Technical data . . . . . . . . . . . . . . . . . . . .|41|
|13.1|Input. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .|41|
|13.2|Output. . . . . . . . . . . . . . . . . . . . . . . . . . . . .|41|
|13.3|Power supply. . . . . . . . . . . . . . . . . . . . . . . . .|41|
|13.4|Performance characteristics. . . . . . . . . . . . . .|41|
|13.5|Installation. . . . . . . . . . . . . . . . . . . . . . . . . .|42|
|13.6|Environment. . . . . . . . . . . . . . . . . . . . . . . . .|42|
|13.7|Process. . . . . . . . . . . . . . . . . . . . . . . . . . . . .|42|
|13.8|Mechanical construction. . . . . . . . . . . . . . . .|43|
|13.9|Operability. . . . . . . . . . . . . . . . . . . . . . . . . .|45|
|13.10|Certificates and approvals. . . . . . . . . . . . . . .|46|
Endress+Hauser
3
Table of contents
Picomag IO-Link
- 14 Appendix . . . . . . . . . . . . . . . . . . . . . . . . . 47 14.1 Radio approvals . . . . . . . . . . . . . . . . . . . . . . . 47 14.1.1 Argentina . . . . . . . . . . . . . . . . . . . . . 47 14.1.2 Brazil . . . . . . . . . . . . . . . . . . . . . . . . 47 14.1.3 Europe . . . . . . . . . . . . . . . . . . . . . . . 47 14.1.4 India . . . . . . . . . . . . . . . . . . . . . . . . . 47 14.1.5 Indonesia . . . . . . . . . . . . . . . . . . . . . 48 14.1.6 Canada and United States . . . . . . . . . 48 14.1.7 Mexico . . . . . . . . . . . . . . . . . . . . . . . 48 14.1.8 Republic of Korea . . . . . . . . . . . . . . . 49 14.1.9 Singapore . . . . . . . . . . . . . . . . . . . . . 49 14.1.10 Taiwan . . . . . . . . . . . . . . . . . . . . . . . 49 14.1.11 Thailand . . . . . . . . . . . . . . . . . . . . . . 49 14.1.12 United Arab Emirates . . . . . . . . . . . . 50 14.1.13 Other countries . . . . . . . . . . . . . . . . . 50
- 14.2 IO-Link process data . . . . . . . . . . . . . . . . . . . 51 14.2.1 Data structure . . . . . . . . . . . . . . . . . . 51 14.2.2 Diagnostic information . . . . . . . . . . . 51
- 14.3 IO-Link ISDU parameter list . . . . . . . . . . . . . . 52 Index . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 59
Endress+Hauser
4
Picomag IO-Link
About this document
## 1 About this document
## 1.1 Document function
These Operating Instructions contain all the information that is required in various phases of the life cycle of the device: from product identification, incoming acceptance and storage, to mounting, connection, operation and commissioning through to troubleshooting, maintenance and disposal.
## 1.2 Symbols used
## 1.2.1 Safety symbols
## **DANGER**
This symbol alerts you to a dangerous situation. Failure to avoid this situation will result in serious or fatal injury.
**WARNING** This symbol alerts you to a dangerous situation. Failure to avoid this situation can result in serious or fatal injury.
## **CAUTION**
This symbol alerts you to a dangerous situation. Failure to avoid this situation can result in minor or medium injury.
## **NOTICE**
This symbol contains information on procedures and other facts which do not result in personal injury.
## 1.2.2 Electrical symbols
|Symbol|Meaning|
|---|---|
||Direct current|
||Ground connection<br>A grounded terminal which, as far as the operator is concerned, is grounded via a<br>grounding system.|
## 1.2.3 Communication-specific symbols
|Symbol|Meaning|
|---|---|
||Bluetooth®<br>Wireless data transmission between devices over a short distance.|
|®|IO-Link<br>Communications system for connecting intelligent sensors and actuators to an<br>automation system. In the IEC 61131-9 standard, IO-Link is standardized under the<br>name "Single-drop digital communication interface for small sensors and actuators<br>(SDCI)".|
Endress+Hauser
5
About this document
Picomag IO-Link
## 1.2.4 Symbols for certain types of information
|Symbol|Meaning|
|---|---|
||Permitted<br>Procedures, processes or actions that are permitted.|
||Preferred<br>Procedures, processes or actions that are preferred.|
||Forbidden<br>Procedures, processes or actions that are forbidden.|
||Tip<br>Indicates additional information.|
||Reference to documentation|
|**A**|Reference to page|
||Reference to graphic|
||Notice or individual step to be observed|
|1.,<br>2.,<br>3.…|Series of steps|
||Result of a step|
## 1.2.5 Symbols in graphics
|Symbol|Meaning|
|---|---|
|1, 2, 3, ...|Item numbers|
|A, B, C, ...|Views|
## 1.3 Documentation
For an overview of the scope of the associated Technical Documentation, refer to the following:
- The _W@M Device Viewer_ : Enter the serial number of the measuring device (www.endress.com/deviceviewer)
- The _Endress+Hauser Operations App_ : Enter the serial number of the measuring device or scan the 2-D matrix code on the measuring device.
## 1.4 Registered trademarks
## ®
Is a registered trademark. It may only be used in conjunction with products and services by members of the IO-Link Community or by non-members who hold an appropriate license. For more specific guidelines on use, refer to the IO-Link Community rules on: www.io.link.com.
## Bluetooth® wireless technology
®
The Bluetooth® word mark and logos are registered trademarks owned by the Bluetooth SIG, Inc. and any use of such marks by Endress+Hauser is under license.
## Apple®
Apple, the Apple logo, iPhone, and iPod touch are trademarks of Apple Inc., registered in the U.S. and other countries. App Store is a service mark of Apple Inc.
Endress+Hauser
6
Picomag IO-Link
About this document
## Android®
Android, Google Play and the Google Play logo are trademarks of Google Inc.
Endress+Hauser
7
Picomag IO-Link
Basic safety instructions
## 2 Basic safety instructions
## 2.1 Requirements for the personnel
The personnel for installation, commissioning, diagnostics and maintenance must fulfill the following requirements:
- ‣ Trained, qualified specialists must have a relevant qualification for this specific function and task.
- ‣ Are authorized by the plant owner/operator.
- ‣ Are familiar with federal/national regulations.
- ‣ Before starting work, read and understand the instructions in the manual and supplementary documentation as well as the certificates (depending on the application).
- ‣ Follow instructions and comply with basic conditions.
The operating personnel must fulfill the following requirements:
- ‣ Are instructed and authorized according to the requirements of the task by the facility's owner-operator.
- ‣ Follow the instructions in this manual.
## 2.2 Intended use
## Application and media
The measuring device described in this manual is intended only for the flow measurement of liquids with a minimum conductivity of 10 µS/cm.
To ensure that the measuring device remains in proper condition for the operation time:
- ‣ Use the measuring device only for media to which the process-wetted materials have an adequate level of resistance.
## Incorrect use
Non-designated use can compromise safety. The manufacturer is not liable for damage caused by improper or non-designated use.
## L[WARNING]
Danger of breakage due to corrosive or abrasive fluids and ambient conditions!
- ‣ Verify the compatibility of the process fluid with the sensor material.
- ‣ Ensure the resistance of all fluid-wetted materials in the process.
- ‣ Keep within the specified pressure and temperature range.
## Residual risks
## L[WARNING]
If the temperature of the media or electronics unit is high or low, this may cause the surfaces of the device to become hot or cold. This poses a risk of burns or frostbite!
- ‣ In the case of hot or cold medium temperatures, install appropriate protection against contact.
## 2.3 Workplace safety
For work on and with the device:
- ‣ Wear the required personal protective equipment according to national regulations.
For welding work on the piping:
- ‣ Do not ground the welding unit via the measuring device.
Endress+Hauser
8
Picomag IO-Link
Basic safety instructions
## 2.4 Operational safety
Risk of injury!
- ‣ Operate the device in proper technical condition and fail-safe condition only.
- ‣ The operator is responsible for interference-free operation of the device.
## 2.5 Product safety
This measuring device is designed in accordance with good engineering practice to meet state-of-the-art safety requirements, has been tested, and left the factory in a condition in which it is safe to operate.
It meets general safety standards and legal requirements. It also complies with the EU directives listed in the device-specific EU Declaration of Conformity. Endress+Hauser confirms this by affixing the CE mark to the device.
Furthermore, the device meets the legal requirements of the applicable UK regulations (Statutory Instruments). These are listed in the UKCA Declaration of Conformity along with the designated standards.
By selecting the order option for UKCA marking, Endress+Hauser confirms a successful evaluation and testing of the device by affixing the UKCA mark.
Contact address Endress+Hauser UK: Endress+Hauser Ltd. Floats Road Manchester M23 9NF United Kingdom www.uk.endress.com
## 2.6 IT security
Our warranty is valid only if the device is installed and used as described in the Operating Instructions. The device is equipped with security mechanisms to protect it against any inadvertent changes to the settings.
IT security measures, which provide additional protection for the device and associated data transfer, must be implemented by the operators themselves in line with their security standards.
## 2.7 Device-specific IT security
## 2.7.1 Access via the SmartBlue app
Two access levels (user roles) are defined for the device: the Operator user role and the Maintenance user role. The Maintenance user role is the default setting.
If a user-specific access code is not defined (in the Set access code parameter), the default setting 0000 continues to apply and the Maintenance user role is automatically enabled. The device's configuration data are not write-protected and can be edited at all times.
If a user-specific access code has been defined (in the Set access code parameter), all the parameters are write-protected and the device is accessed with the Operator user role. The previously defined access code must first be entered again before the Maintenance user role is enabled and all the parameters can be write-accessed.
## 2.7.2 Protecting access via a password
Different passwords are available to protect write access to the device parameters or access to the device via the Bluetooth® interface.
Endress+Hauser
9
Picomag IO-Link
Basic safety instructions
- User-specific access code
- Protect write access to the device parameters via the SmartBlue app
- Bluetooth key
- The password protects a connection between an operating device (e.g. smartphone, tablet) and the device via the Bluetooth® interface.
General notes on the use of passwords
- The access code and Bluetooth key supplied with the device should be changed during commissioning.
- Follow the general rules for generating a secure password when defining and managing the access code or Bluetooth key.
- The user is responsible for the management and careful handling of the access code and Bluetooth key.
## 2.7.3 Access via Bluetooth® wireless technology
Secure signal transmission via Bluetooth® wireless technology uses an encryption method tested by the Fraunhofer Institute.
- The device is not visible via _Bluetooth_ ® wireless technology without the SmartBlue App.
- Only one point-to-point connection is established between the device and a smartphone or tablet.
- The _Bluetooth_ ® wireless technology interface can be disabled via SmartBlue.
Endress+Hauser
10
Picomag IO-Link
Incoming acceptance and product identification
- 3 Incoming acceptance and product identification
## 3.1 Incoming acceptance
**==> picture [287 x 56] intentionally omitted <==**
**----- Start of picture text -----**<br>
Are the order codes on the<br>delivery note (1) and the<br>1<br>product sticker (2) identical?<br>2<br>**----- End of picture text -----**<br>
**==> picture [121 x 86] intentionally omitted <==**
**==> picture [349 x 229] intentionally omitted <==**
**----- Start of picture text -----**<br>
Are the goods undamaged?<br>Do the data on the<br>Ext. ord. cd.:Order code:Ser. no.: nameplate match theordering information on the<br>delivery note?<br>i i<br>Date:<br>Is the accompanying safety<br>data sheet present?<br>**----- End of picture text -----**<br>
- If one of the conditions is not satisfied, contact your Endress+Hauser Sales Center.
- The Technical Documentation is available via the Internet or via the _Endress+Hauser Operations App_ , see the "Product identification" section .
## 3.2 Product identification
The following options are available for identification of the measuring device:
- The device label
- Order code with breakdown of the device features on the delivery note
- Enter the serial number on the device label in _W@M Device Viewer_
- (www.endress.com/deviceviewer): all the information about the measuring device is displayed.
- Enter the serial number on the device label into the _Endress+Hauser Operations App_ or scan the 2-D matrix code (QR code) on the measuring device with the _Endress+Hauser Operations App_ : all the information about the measuring device is displayed.
Endress+Hauser
11
Incoming acceptance and product identification
Picomag IO-Link
## 3.2.1 Symbols on measuring device
|Symbol|Meaning|
|---|---|
||WARNING!<br>This symbol alerts you to a dangerous situation. Failure to avoid this situation can result in serious<br>or fatal injury. To determine the nature of the potential hazard and the measures required to<br>avoid it, consult the documentation accompanying the measuring device.|
||Reference to documentation<br>Refers to the corresponding device documentation.|
Endress+Hauser
12
Picomag IO-Link
Storage and transport
## 4 Storage and transport
## 4.1 Storage conditions
Observe the following notes for storage:
- ‣ Store in the original packaging to ensure protection from shock.
- ‣ Store in a dry place.
- ‣ Do not store outdoors.
Storage temperature → 42
## 4.2 Transporting the product
Transport the device to the measuring point in the original packaging.
Do not remove protective covers or caps installed on process connections. They prevent mechanical damage to the sealing surfaces and contamination in the measuring tube.
## 4.3 Packaging disposal
All packaging materials are environmentally friendly and 100% recyclable: Carton in accordance with European Packaging Directive 94/62EC; recyclability is confirmed by the affixed RESY symbol.
Endress+Hauser
13
Picomag IO-Link
Mounting
## 5 Mounting
## 5.1 Mounting requirements
## 5.1.1 Mounting position
## Mounting location
**==> picture [239 x 119] intentionally omitted <==**
**==> picture [22 x 4] intentionally omitted <==**
**----- Start of picture text -----**<br>
A0046065<br>**----- End of picture text -----**<br>
Preferably install the sensor in an ascending pipe.
## Inlet and outlet runs
No inlet and outlet runs need to be considered.
- The installation dimensions provide information on the dimensions and installed lengths of the device → 43
- The arrow points in the preferred direction of flow. Measurement in the other direction is also possible.→ 23
## 5.2 Mounting the measuring device
Install the measuring device without any mechanical tension so that pipe forces are not transferred to the measuring device.
## L[WARNING]
## Burn hazard!
If medium temperatures or ambient temperatures exceed 50 °C, areas of the housing can heat to over 65 °C.
- ‣ Safeguard the housing so that it cannot be touched accidentally.
Endress+Hauser
14
Picomag IO-Link
Mounting
**==> picture [387 x 162] intentionally omitted <==**
**----- Start of picture text -----**<br>
1 2 3 4 5 5 4 3 2 1<br> A0033002<br> 1 Measuring device with male thread<br>1 Pipe with female thread<br>2 Seal (not supplied)<br>3 Adapter: available adapters → 39<br>4 Seal (included in delivery)<br>5 Measuring device connection, male thread<br>**----- End of picture text -----**<br>
**==> picture [387 x 132] intentionally omitted <==**
**----- Start of picture text -----**<br>
1 2 2 1<br> A0046929<br> 2 Measuring device with female thread<br>1 Pipe with conical male thread<br>2 Measuring device connection, female thread<br>**----- End of picture text -----**<br>
No seals are included in the delivery for devices with a female thread. Suitable commercially available seal material can be used.
Endress+Hauser
15
Electrical connection
Picomag IO-Link
## 6 Electrical connection
## 6.1 Electrical safety
In accordance with applicable national regulations.
## 6.2 Connecting requirements
## 6.2.1 Requirements for connecting cable
National regulations and standards apply.
|Connecting cable|M12 × 1 A-coded|
|---|---|
|Conductor cross-section|At least 0.12 mm2(AWG26)|
|Degree of protection|IP65/67, pollution degree 3|
## 6.2.2 Pin assignment, device plug
**==> picture [169 x 236] intentionally omitted <==**
**----- Start of picture text -----**<br>
A<br>B<br>1 3<br>2 4 2 4<br>3 1<br>A B<br>1 L+<br>2 IO2 �<br>4 IO1<br>3 L-<br>+<br>-<br>**----- End of picture text -----**<br>
**==> picture [22 x 4] intentionally omitted <==**
**----- Start of picture text -----**<br>
A0033004<br>**----- End of picture text -----**<br>
- _A Plug (measuring device)_
- _B Socket (customer side)_
- _1...4 Pin assignment_
## _Pin assignment_
|Pin|Assignment|Description|
|---|---|---|
|1|L+|Supply voltage + (18 to 30 VDC/max. 3 W)|
|2|IO2|Input/output 2, can be configured independently of IO1|
|3|L-|Supply voltage -|
|4|IO1|Input/output 1, can be configured independently of IO2|
Endress+Hauser
16
Picomag IO-Link
Electrical connection
## Switch output configuration version
The switching behavior of IO1 and IO2 can be configured independently of one another.
**==> picture [387 x 190] intentionally omitted <==**
**----- Start of picture text -----**<br>
PNP [ 1)] NPN [ 2)]<br>A B A B<br>1 L+ 1 L+<br>2 2<br>4 4<br>3 L- 3 L-<br> A0033005 A0033006<br>A Plug (measuring device) A Plug (measuring device)<br>B Socket (customer side) B Socket (customer side)<br>L+ Supply voltage + L+ Supply voltage +<br>L- Supply voltage - L- Supply voltage -<br>The load is switched to high side L+. The maximum The load is switched to low side L-. The maximum load<br>load current is 250 mA. The output is protected current is 250 mA. The output is protected against<br>against short-circuiting and reverse polarity. short-circuiting and reverse polarity.<br>**----- End of picture text -----**<br>
- 1) Positive Negative Positive (high side switch)
- 2) Negative Positive Negative (low side switch)
## Pulse output configuration version
Option only available for output 1 → 25
**==> picture [112 x 67] intentionally omitted <==**
**----- Start of picture text -----**<br>
A B<br>1 L+<br>2<br>4<br>3 L-<br>**----- End of picture text -----**<br>
- A0047257
- _ 3 Pulse output with PNP switching behavior_
- _A Plug (measuring device)_
- _B Socket (customer side)_
- _L+ Supply voltage +_
- _LSupply voltage -_
The load is switched to high side L+. The maximum load current is 250 mA. The output is protected against short-circuiting and reverse polarity.
Endress+Hauser
17
Electrical connection
Picomag IO-Link
## Current output configuration version
**==> picture [387 x 97] intentionally omitted <==**
**----- Start of picture text -----**<br>
A B<br>1 L+<br>2<br>4<br>A A<br>3 L-<br> A0046581<br>**----- End of picture text -----**<br>
- _ 4 Current output, active, 4 to 20 mA_
- _A Plug (measuring device)_
- _B Socket (customer side)_
- _L+ Supply voltage +_
- _LSupply voltage -_
The current flows from the output to L-. The maximum load may not exceed 500 Ω. A bigger load distorts the output signal.
## Voltage output configuration version
**==> picture [387 x 98] intentionally omitted <==**
**----- Start of picture text -----**<br>
A B<br>1 L+<br>2<br>4<br>V V<br>3 L-<br> A0046582<br>**----- End of picture text -----**<br>
- _ 5 Voltage output, active, 2 to 10 V_
- _A Plug (measuring device)_
- _B Socket (customer side)_
- _L+ Supply voltage +_
- _LSupply voltage -_
The voltage from the output applies to L-. The load must be at least 600 Ω. The output is protected against short-circuiting and reverse polarity.
## Digital input configuration version
- 15 V (switch-on threshold)
- 5 V (switch-off threshold)
**==> picture [387 x 97] intentionally omitted <==**
**----- Start of picture text -----**<br>
A B<br>1 L+<br>2<br>4<br>= =<br>3 L-<br> A0033008<br>**----- End of picture text -----**<br>
- _ 6 Digital input_
- _A Plug (measuring device)_
- _B Socket (customer side)_
- _L+ Supply voltage +_
- _LSupply voltage -_
Endress+Hauser
18
Picomag IO-Link
Electrical connection
A0046190
Internal resistance: 7.5 kΩ
## IO-Link configuration version
Option only available for output 1 in the Output 1→ 24 submenu
The measuring device features an IO-Link communication interface with a baud rate of 38,400 and with a second IO function on pin 2. This requires an IO-Link compatible module (IO-Link master) for operation. The IO-Link communication interface allows direct access to the process and diagnostics data.
## 6.3 Connecting the measuring device
## NOTICE
The measuring device may only be installed by properly trained technicians.
- ‣ Comply with national and international regulations regarding the installation of electrotechnical systems.
- ‣ Power supply according to EN 50178, SELV, PELV or Class 2.
1. De-energize the system.
2. Connect the measuring device via the connector.
3.
**==> picture [34 x 128] intentionally omitted <==**
**==> picture [6 x 7] intentionally omitted <==**
**----- Start of picture text -----**<br>
1<br>**----- End of picture text -----**<br>
- _1 Ground terminal_
In the case of non-grounded pipes:
The device must be grounded using the ground terminal accessory.
## 6.4 Post-connection check
|6.4<br>Post-connection check||
|---|---|
|Are cables or the device undamaged (visual inspection)?||
|Do the cables have adequate strain relief?||
|Is the connector connected correctly?||
|Does the supply voltage match the specifications on the measuring device?||
|Is the pin assignment of the connector correct?||
|Is the potential equalization established correctly?||
Endress+Hauser
19
Picomag IO-Link
Operation options
## 7 Operation options
## 7.1 Access to the operating menu via the SmartBlue app
The device can be operated and configured via the SmartBlue app. In this case, the connection is established via the Bluetooth® wireless technology interface.
## _Supported functions_
- Device selection in Live List and access to the device (login)
- Configuration of the device
- Access to measured values, device status and diagnostic information
The SmartBlue app is available to download free of charge for Android devices (Google Playstore) and iOS devices (iTunes Apple Store): _Endress+Hauser SmartBlue_
Directly to the app with the QR code:
**==> picture [22 x 4] intentionally omitted <==**
**----- Start of picture text -----**<br>
A0033202<br>**----- End of picture text -----**<br>
## System requirements
- Devices with iOS:
- From iOS 11
- Devices with Android:
- Android 6.0 KitKat or higher
Downloading the SmartBlue app:
1. Install and start the SmartBlue app.
- A Live List shows all the devices available.
- The list displays the devices with the configured tag name. The default setting for the tag name is EH_DMA_XYZZ (XYZZ = the last 7 digits of the device serial number).
2. For Android devices, activate GPS positioning (not necessary for devices with IOS)
3. Select the device from the Live List.
- The Login dialog box opens.
Logging in:
4. Enter the user name: admin
5. Enter the initial password: serial number of the device.
- When you log in for the first time, a message is displayed advising you to change the password.
6. Confirm your entry.
- The main menu opens.
7. Optional: Change Bluetooth® password: Menu: Guidance → Security → Change Bluetooth password
Forgotten your password: contact Endress+Hauser Service.
- Navigate through the various items of information about the device: swipe the screen to the side.
Endress+Hauser
20
Picomag IO-Link
System integration
## 8 System integration
The measuring device has an IO-Link communication interface. The IO-Link interface allows direct access to process and diagnostics data and enables the user to configure the measuring device on the fly.
Properties:
- IO-Link Specification: Version 1.1
- IO-Link Smart Sensor Profile 2nd Edition
- SIO mode: yes
- Speed: COM2 (38.4 kBaud)
- Minimum cycle time: 10 ms
- Process data width: 120 bit
- IO-Link data storage: yes
- Block configuration: no
- Device operational: The measuring device is operational 4 seconds after the supply voltage is applied
More information on IO-Link is available at www.io-link.com
Overview of the entire IO-Link ISDU parameter list → 52
## 8.1 Overview of device description files
_Current version data for the device_
|_Current version data for_|_the device_||
|---|---|---|
|Firmware version|01.01.zz|Information on the firmware version can be found:<br>• On the title page of the Operating instructions<br>• On the device label<br>• In theFirmware versionparameter in the System menu|
|Release date of firmware<br>version|05.2019|---|
|Profile version|• 1.1<br>• Smart Sensor<br>Profile|---|
## 8.2 Device master file
In order to integrate field devices into a digital communication system, the IO-Link system needs a description of the device parameters, such as output data, input data, data format, data volume and supported transfer rate.
These data are available in the device master file (IODD[ 1)] ), provided to the IO-Link master when the communication system is commissioned.
The IODD can be downloaded as follows:
- Endress+Hauser: www.endress.com
- IODDfinder: http://ioddfinder.io-link.com
1) IO Device Description
Endress+Hauser
21
Picomag IO-Link
Commissioning
## 9 Commissioning
## 9.1 Switching ON the measuring device
Once the supply voltage has been switched on, the measuring device adopts the normal mode after a maximum of 4 s. During the start-up phase, the outputs are in the same state as the measuring device in the switched-off state.
## 9.2 Overview of the operating menu
## _Overview of the operating menu_
**==> picture [357 x 309] intentionally omitted <==**
**----- Start of picture text -----**<br>
Guidance<br>▸ Identification → 22<br>▸ System units → 22<br>▸ Sensor → 23<br>▸ Output 1 → 24<br>▸ Output 2 → 24<br>▸ Totalizer → 29<br>▸ Display → 30<br>▸ Data management → 31<br>▸ Security → 31<br>Diagnostics → 32<br>System<br>▸ Data management → 32<br>▸ Firmware → 32<br>**----- End of picture text -----**<br>
## 9.3 Configuring the measuring device
## 9.3.1 Identification
The device tag and the user level can be changed in the Identification submenu. Navigation
Menu: "Guidance" → Identification
## 9.3.2 Configuring system units
In the System units submenu, you can configure the units of all measured values. Navigation
Endress+Hauser
22
Picomag IO-Link
Commissioning
Menu: "Guidance" → System units
_Parameter overview with brief description_
|Parameter|Description|Selection|Factory setting|
|---|---|---|---|
|Volume flow unit|Select the unit for the volume flow.|• l/s, m³/h, l/min, l/h<br>• gal/min (us), fl. oz/min|l/min|
|Volume unit|Select the unit for the volume.|• ml, l, m³<br>• fl. oz (us), gal (us)|ml|
|Temperature unit|Select the unit for the temperature.|• °C<br>• °F|°C|
|Totalizer unit|Select the unit for the totalizer.|• ml, l, m³<br>• kl, Ml<br>• fl. oz (us), gal (us)<br>• kgal (us)|m³|
|Conductivity unit|Select the unit for conductivity.|• µS/cm<br>• S/m<br>• mS/cm|µS/cm|
## 9.3.3 Setting the installation direction and measurement
The Sensor submenu contains parameters for specific settings of the measuring device. Navigation
Menu: "Guidance" → Sensor
## _Parameter overview with brief description_
|Parameter|Description|Selection/input|Factory setting|
|---|---|---|---|
|Installation<br>direction|Select the installation<br>direction.|• Flow in arrow direction (forwards)<br>Positive flow measurement in the direction of the arrow.<br>• Flow against arrow direction (backwards)<br>Positive flow measurement in the opposite direction of the arrow.|Flow in arrow direction<br>(forwards)|
|On value|Enter the on value for<br>low flow cut off.|Positive floating point number<br>A flow measured value that is less than the value of the on value<br>forces the display to zero. In the event of plant downtime, this<br>prevents the totalizer from continuing to totalize even though<br>there is no flow.<br>The measured value flashes on the display when low flow cut off is<br>active.<br>Low flow cut off is deactivated again as soon as the flow measured<br>value exceeds 1.5 times the on value.|Depends on the nominal<br>diameter:<br>DN 15 (½"):<br>0.05 l/min (0.013 gal/min)<br>DN 20 (¾"):<br>0.1 l/min (0.026 gal/min)<br>DN 25 (1"):<br>0.2 l/min (0.052 gal/min)<br>DN 50 (2"):<br>1.5 l/min (0.4 gal/min)|
|Damping|Enter the time constant<br>for damping the flow<br>measured value.|0 to 10 s|0 s|
Endress+Hauser
23
Picomag IO-Link
Commissioning
## 9.3.4 Configuring the IO modules
The measuring device has two signal inputs or signal outputs that can be configured independently of one another:
- Current output → 24
- Pulse output → 25
- Switch output → 26
- Voltage output → 27
- Digital input → 28
## Navigation
Menu: "Guidance" → Output 1
Menu: "Guidance" → Output 2
## _Parameter overview with brief description_
|Parameter|Description|Selection|Factory setting|
|---|---|---|---|
|Output 1|Select the operating mode of output 1.|• Pulse output<br>• Current output<br>• Switch output<br>• Voltage output<br>• Digital input<br>• IO-Link<br>• Off|IO-Link|
|Output 2|Select the operating mode of output 2.|• Current output<br>• Switch output<br>• Voltage output<br>• Digital input<br>• Off|Off|
## Configuring the current output
The Current output submenu contains all the parameters that must be configured for the configuration of the current output.
The output is used to output process variables by analog means in the form of a 4-20 mA current.
## Navigation
Menu: "Guidance" → Output 1 → Current output
Menu: "Guidance" → Output 2 → Current output
## _Parameter overview with brief description_
|Parameter|Description|Selection/input|Factory setting|
|---|---|---|---|
|Assign current output|Select process variable for current output.|• Off<br>• Volume flow<br>• Temperature<br>• Conductivity|Volume flow|
|4 mA value|Enter 4 mA value.|Floating point number with sign|0 l/min|
|20 mA value|Enter 20 mA value.|Floating point number with sign|Depends on the nominal diameter:<br>DN 15 (½"): 25 l/min (6.6 gal/min)<br>DN 20 (¾"): 50 l/min (13.2 gal/min)<br>DN 25 (1"): 100 l/min (26.4 gal/min)<br>DN 50 (2"): 750 l/min (198.1 gal/min)|
Endress+Hauser
24
Picomag IO-Link
Commissioning
_Unidirectional flow measurement (Q), conductivity measurement (S)_
**==> picture [387 x 169] intentionally omitted <==**
**----- Start of picture text -----**<br>
mA<br>20<br>16<br>12<br>8<br>4<br>Q<br>A = 0 B<br> A0035753<br>A Lower range value = 0<br>B Upper range value<br>Q Flow<br>**----- End of picture text -----**<br>
- Current I is linearly interpolated between lower range value (A) and upper range value (B).
- The output range ends at 20.5 mA.
_Bidirectional flow measurement (Q), temperature measurement (T)_
**==> picture [387 x 170] intentionally omitted <==**
**----- Start of picture text -----**<br>
mA<br>20<br>16<br>12<br>8<br>4<br>Q<br>A = 0 B<br> A0035754<br>A Lower range value<br>B Upper range value<br>Q Flow<br>**----- End of picture text -----**<br>
- Current I is linearly interpolated between lower range value (A) and upper range value (B).
- Rather than having a hard upper and lower limit, the output range ends at 20.5 mA at the top end and at 3.8 mA at the bottom end
## Configuring the pulse output
The Pulse output submenu contains all the parameters that must be configured for the configuration of the pulse output.
## Navigation
Menu: "Guidance" → Output 1
## _Parameter overview with brief description_
|Parameter|Description|User entry|Factory setting|
|---|---|---|---|
|Value per pulse|Enter the value for the pulse output.|Floating point number with sign|Depends on the nominal diameter:<br>DN 15 (½"): 0.5 ml<br>DN 20 (¾"): 1.0 ml<br>DN 25 (1"): 2.0 ml<br>DN 50 (2"): 10.0 ml|
Endress+Hauser
25
Picomag IO-Link
Commissioning
The current pulse repetition frequency is calculated from the current flow and the configured pulse value:
Pulse repetition frequency = flow/pulse value
## Example
- Flow: 24 l/min
- Pulse value: 0.001 l
- Pulse repetition frequency = 400 Pulse/s
The pulse output only outputs positive flow components in the set installation direction. Negative flow components are ignored and not balanced.
Low flow cut off → 23 resets the pulse output.
## Configuring the switch output
The Switch output submenu contains all the parameters that must be configured for the configuration of the switch output.
## Navigation
Menu: "Guidance" → Output 1
Menu: "Guidance" → Output 2
## _Parameter overview with brief description_
|Parameter|Description|Selection/input|Factory setting|
|---|---|---|---|
|Polarity|Select the switching<br>behavior.|• NPN (low-side-switch)<br>Switches load to low side to L-<br>• PNP (high-side-switch)<br>Switches load to high side to L+|PNP (high-side-<br>switch)|
|Switch output<br>function||• Off<br>The switch output is permanently switched off (open, non-conductive).<br>• On<br>The switch output is permanently switched on (closed, conductive).<br>• Diagnostic behavior<br>The output switches off when an event with the status signal F occurs<br>• Limit volume flow<br>Indicates if a specified limit value has been reached for the process variable.<br>• Limit temperature<br>Indicates if a specified limit value has been reached for the process variable.<br>• Limit conductivity<br>Indicates if a specified limit value has been reached for the process variable.<br>• Limit volume totalizer<br>• Range volume flow<br>• Range temperature<br>• Range conductivity<br>• Range volume totalizer<br>• Empty pipe detection<br>Output switches off if empty pipe detection is activated.|Off|
|Switch-on value|Enter the measured value<br>for the switch-on value.|Floating point number with sign|1 000 m³/h|
|Switch-off value|Enter the measured value<br>for the switch-off value.|Floating point number with sign|1 000 m³/h|
Endress+Hauser
26
Picomag IO-Link
Commissioning
**==> picture [387 x 440] intentionally omitted <==**
**----- Start of picture text -----**<br>
Switch on if over the limit: Switch off if over the limit:<br>Q / T Q / T<br>C C<br>A B<br>B A<br>t t<br>1 1<br>0 0<br> A0034163 A0034164<br>1.1 Input variables 2.1 Input variables<br>1.2 Switch output 2.2 Switch output<br>A Switch-on point A Switch-on point<br>B Switch-off point B Switch-off point<br>C Hysteresis C Hysteresis<br>Switch on if within window: Switch off if within window:<br>Q / T Q / T<br>D D D D<br>B B A A<br>A A B B<br>t t<br>1 1<br>0 0<br> A0034165 A0034166<br>3.1 Input variables 4.1 Input variables<br>3.2 Switch output 4.2 Switch output<br>A On-value (lower range limit) A On-value (lower range limit)<br>B Off-value (upper range limit) B Off-value (upper range limit)<br>D Window D Window<br>1.1 2.1<br>1.2 2.2<br>3.1 4.1<br>3.2 4.2<br>**----- End of picture text -----**<br>
## Configuring the voltage output
The Voltage output submenu contains all the parameters that must be configured for the configuration of the voltage output.
## Navigation
Menu: "Guidance" → Output 1
Menu: "Guidance" → Output 2
## _Parameter overview with brief description_
|Parameter|Description|Selection/input|Factory setting|
|---|---|---|---|
|Assign voltage output|Select process variable for voltage output.|• Off<br>• Volume flow<br>• Temperature<br>• Conductivity|Volume flow|
|2 V value|Enter the lower range value.|Floating point number with sign|0 l/min|
|10 V value|Enter the upper range value.|Floating point number with sign|Depends on the nominal diameter:<br>DN 15 (½"): 25 l/min<br>DN 20 (¾"): 50 l/min<br>DN 25 (1"): 100 l/min<br>DN 50 (2"): 750 l/min|
Endress+Hauser
27
Picomag IO-Link
Commissioning
_Unidirectional flow measurement (Q), conductivity measurement_
**==> picture [229 x 113] intentionally omitted <==**
**----- Start of picture text -----**<br>
V<br>10<br>8<br>6<br>4<br>2<br>Q<br>A = 0 B<br>**----- End of picture text -----**<br>
**==> picture [22 x 4] intentionally omitted <==**
**----- Start of picture text -----**<br>
A0032995<br>**----- End of picture text -----**<br>
- _A Lower range value = 0 B Upper range value Q Flow_
- Voltage U is linearly interpolated between lower range value (A) and upper range value (B).
- The output range ends at 10.25 V.
_Bidirectional flow measurement (Q), temperature measurement (T)_
**==> picture [232 x 116] intentionally omitted <==**
**----- Start of picture text -----**<br>
V<br>10<br>8<br>6<br>4<br>2<br>Q<br>A 0 B<br>**----- End of picture text -----**<br>
**==> picture [22 x 4] intentionally omitted <==**
**----- Start of picture text -----**<br>
A0032996<br>**----- End of picture text -----**<br>
- _A Lower range value B Upper range value Q Flow_
- Voltage U is linearly interpolated between lower range value (A) and upper range value (B).
- Rather than having a hard upper and lower limit, the output range ends at 10.25 V at the top end and at 1.9 V at the bottom end
## Configuring the digital input
The Digital input submenu contains all the parameters that must be configured for the configuration of the digital input.
The input is used to control an action with an external voltage signal. The minimum pulse duration is 100 ms.
## Navigation
Menu: "Guidance" → Output 1
Endress+Hauser
28
Picomag IO-Link
Commissioning
## Menu: "Guidance" → Output 2
_Parameter overview with brief description_
|Parameter|Description|Selection|Factory setting|
|---|---|---|---|
|Active level|Select the switching behavior for the digital<br>input.|• High<br>Input reacts to high level/positive edge<br>• Low<br>Input reacts to low level/negative edge|High|
|Assign status input|Select the function for the digital input.|• Off<br>• Reset totalizer<br>Reset the totalizer (edge controlled)<br>(positive/negative edge)<br>• Flow override<br>• Flow measured value = 0<br>• No effect on temperature measurement (level controlled)<br>(high/low level)|Reset totalizer|
## 9.3.5 Totalizer
The totalizer can be reset with the Reset totalizer submenu.
## Navigation
Menu: "Guidance" → Totalizer
_Parameter overview with brief description_
|Parameter|Description|Display/options|Factory setting|
|---|---|---|---|
|Volume totalizer|Enter value.|Floating point number with sign|0 m3|
|Reset totalizer|Reset the totalizer.|• Cancel<br>The totalizer is not reset.<br>• Reset + totalize<br>The totalizer is reset.|Cancel|
Endress+Hauser
29
Picomag IO-Link
Commissioning
## 9.3.6 Configuring the display
The Display submenu contains all the parameters that can be configured for the configuration of the onsite display.
## Navigation
Menu: "Guidance" → Display
## _Parameter overview with brief description_
|Parameter|Description|Selection/input|Factory setting|
|---|---|---|---|
|Format display|Select how measured values are<br>shown on the display.|Display value 1st line + display value 2nd line:<br>• Volume flow + temperature<br>• Volume flow + totalizer<br>• Temperature + totalizer<br>• Volume flow + conductivity<br>• Totalizer + conductivity<br>• Temperature + conductivity|Volume flow +<br>temperature|
|||4 display values:<br>Volume flow + temperature + totalizer + conductivity||
|||2 display values (multiplex):<br>Volume flow +totalizer / temperature + conductivity||
|Rotation display|Select local display rotation.|• Auto (automatic)<br>• The display rotates automatically depending on the installation position|Auto|
|||• 0°<br>• Can be read in the horizontal installation position with flow from left to<br>right<br>A<br>A0033013||
|||• 90°<br>• Can be read in the vertical installation position with flow from bottom to<br>top<br>A<br>A0033014||
|||• 180°<br>• Can be read in the horizontal installation position with flow from right to<br>left<br>A<br>A0033015||
Endress+Hauser
30
Picomag IO-Link
Commissioning
**==> picture [517 x 169] intentionally omitted <==**
**----- Start of picture text -----**<br>
Parameter Description Selection/input Factory setting<br>• 270°<br>• Can be read in the vertical installation position with flow from top to<br>bottom<br>A<br> A0033016<br>Backlight Set the intensity of the 0 to 100 % 50 %<br>backlighting.<br>**----- End of picture text -----**<br>
## 9.3.7 Data management
## Export configuration as report
The device configuration can be exported as a PDF report and saved in the mobile terminal or forwarded with this function.
## Save configuration to file
The device configuration is saved in the app. The saved device configuration can be transferred to another measuring device using the System → "Load configuration from app" function.
## Navigation
Menu: "Guidance" → Data management
## 9.3.8 Security
The Security submenu contains all the parameters needed to define the access code and configure the Bluetooth connection.
## Navigation
Menu: "Guidance" → Security
_Parameter overview with brief description_
|Parameter|Description|Entry/selection/display|Factory<br>setting|
|---|---|---|---|
|Define access code|Define a user-specific access code to restrict write access to<br>parameters.|4-digit string of numbers|0000|
|Bluetooth|Enable or disable the_Bluetooth_® wireless technology interface.<br>If the interface is disabled, it can only be re-enabled by<br>tapping the device.|• Disable<br>• Disable the interface.<br>• The connection to the measuring<br>device is torn down.<br>• Enable|Enable|
|Change Bluetooth<br>password|Change Bluetooth password<br>Keep the password safe. If it is lost, Endress+Hauser cannot<br>restore access to the device.|Character string comprising numbers,<br>letters and special characters|Device serial<br>number|
## Enable Bluetooth by tapping the device
1. Activate Bluetooth by tapping on the housing three times.
2. Establish a connection to the device via the SmartBlue app.
Endress+Hauser
31
Picomag IO-Link
Commissioning
## 9.4 Diagnostics
The Diagnostics menu enables you to simulate, without a real flow situation, various process variables in the process and the device alarm mode and to verify downstream signal chains (switching of valves or closed-control loops).
## Navigation
"Diagnostics" menu
_Parameter overview with brief description_
|Parameter|Description|Selection/input|Factory setting|
|---|---|---|---|
|Actual diagnostic|A diagnostic event has occurred.|Displays the current diagnostic event along with the diagnostic<br>information.<br>If two or more messages occur simultaneously, the message<br>with the highest priority is shown on the display.|Symbol for<br>diagnostic behavior,<br>diagnostic code and<br>short message.|
|Simulation process<br>variable|Activate the simulation of process<br>variables.|• Off<br>Simulation is deactivated.<br>• On<br>Simulation is activated.<br>Deactivate the simulation again once the test has been<br>performed.|Off|
|Volume flow value|Enter the value for volume flow<br>simulation.|Positive floating point number|–|
|Temperature value|Enter the value for temperature<br>simulation.|Positive floating point number|–|
|Conductivity value|Enter the value for conductivity<br>simulation|Positive floating point number|–|
## 9.5 System
The System submenu contains all the parameters that can be used for the administration of the device.
## Navigation
Menu: "System"
## _Parameter overview with brief description_
|Parameter|Description|Entry/selection/display|Factory setting|
|---|---|---|---|
|Access status tooling|Displays the access status. No user entry/selection possible.|• Operator<br>• Maintenance|Maintenance|
|Enter access code|Enter the defined access code. → 31<br>Restrict write access to parameters in order to protect the device<br>configuration from unauthorized modification.|4-digit string of numbers|0000|
|Device reset|Reset the entire device configuration or some of the configuration to a<br>defined state.|• Cancel<br>• To factory defaults<br>• Restart device|Cancel|
|Export configuration as<br>report|The device configuration can be exported as a PDF report and saved in the<br>mobile device or forwarded with this function.|-|-|
|Save configuration to file|The device configuration is saved in the app. The saved device configuration<br>can be transferred to another measuring device using the System → "Load<br>configuration from file" function.|-|-|
|Load Configuration from<br>file|The saved device configuration can be uploaded to a new measuring device<br>with this function.|-|-|
Endress+Hauser
32
Picomag IO-Link
Operation
## 10 Operation
## 10.1 Offline quick view of configuration
If you tap the top of the housing (e.g. on the arrow indicating the flow direction) with your fist or an object), the device displays an overview of the preset parameters.
**==> picture [155 x 121] intentionally omitted <==**
**----- Start of picture text -----**<br>
4 5 6 7<br>Xxxxx<br>I01: / 8<br>/ PIN4<br>1<br>I02: / 9<br>/ PIN2<br>2 BT: / 11<br>SNo: 12<br>3<br>SW: 13<br>10 14<br>**----- End of picture text -----**<br>
**==> picture [22 x 4] intentionally omitted <==**
**----- Start of picture text -----**<br>
A0036432<br>**----- End of picture text -----**<br>
- _ 7 Information in the status layout, overview of the preconfigured parameters_
- _1 I/O area_
- _2 Bluetooth area 3 Identification area 4 I/O type 1 5 I/O current value type 1_
- _6 I/O type 2 7 I/O current value type 2 8 I/O assignment type 1_
- _9 I/O assignment type 2_
- _10 Bluetooth module status_
- _11 Bluetooth connection status_
- _12 Serial number_
- _13 User role: Operator (O), Maintenance (M) 14 Software version_
## _I/O area (in brackets: item number → 7, 33)_
|I/O type ( 4, 6)|I/O assignment (8, 9)|I/O current value (5, 7)|
|---|---|---|
|S-Out|• Alrt<br>• LimQ<br>• LimT<br>• LimV<br>• Lims<br>• WinQ<br>• WinT<br>• WinV<br>• Wins<br>• EPD<br>• Off<br>• On|• PNPOn<br>• PNPOff<br>• NPNOn<br>• NPNOff|
|I-Out|• s<br>• Q<br>• T<br>Off|xx.x mA|
|U-Out|• s<br>• Q<br>• T<br>Off|xx.x V|
|S-In|• RsT<br>• Ovrd<br>Off|Low<br>High|
|P-Out|Q|PNPOn<br>PNPOff|
|IO-L|PD|• Dis.<br>• Start<br>• Preop.<br>• Oper.|
|Off|–|–|
Endress+Hauser
33
Picomag IO-Link
Operation
_Bluetooth area (in brackets: item number → 7, 33)_
|Bluetooth module status (10)|Bluetooth connection status (11)|
|---|---|
|On|Dis./Con.|
|Off|Dis.|
Endress+Hauser
34
Picomag IO-Link
Diagnostics and troubleshooting
## 11 Diagnostics and troubleshooting
## 11.1 General troubleshooting
## _For local display_
|Error|Possible causes|Remedial action|
|---|---|---|
|Local display dark and no output<br>signals|Supply voltage does not match the<br>voltage specified on the nameplate.|Apply the correct supply voltage<br>→ 41.|
||Supply voltage has incorrect<br>polarity.|Correct the polarity.|
||The connecting cables are not<br>connected correctly.|Check the cable connection and<br>correct if necessary.|
## _For output signals_
|Error|Possible causes|Remedial action|
|---|---|---|
|Device shows correct value on local<br>display, but signal output is<br>incorrect, though in the valid range.|Parametrization errors|Check parameterization and correct<br>it.|
|Device measures incorrectly.|Configuration error or device is<br>operated outside the application.|1. Check and correct parameter<br>configuration.<br>2. Observe limit values specified in<br>the "Technical Data".|
## _For access_
|Error|Possible causes|Remedial action|
|---|---|---|
|Measuring device not in<br>smartphone or tablet live list|Bluetooth communication is<br>disabled|1. Check whether the Bluetooth<br>logo is visible on the local display or<br>not.<br>2. Re-enable Bluetooth<br>communication by tapping the<br>device three times.|
|Device not responding via<br>SmartBlue app|No Bluetooth connection|Enable Bluetooth function on<br>smartphone or tablet.|
||The device is already connected to<br>another smartphone/tablet.|Check whether the device is already<br>connected.|
|Login via SmartBlue app not<br>possible|Device is being put into operation<br>for the first time|Enter initial password (device serial<br>number) and change.|
|Device cannot be operated via<br>SmartBlue app|Incorrect password entered|Enter correct password.|
||Password forgotten|Contact Endress+Hauser Service.|
|No write access to parameters|Current user role has limited access<br>authorization|• Check user role<br>• Enter the correct customer-<br>specific access code → 20|
Endress+Hauser
35
Diagnostics and troubleshooting
Picomag IO-Link
## 11.2 Diagnostic information on local display
## 11.2.1 Diagnostic message
Faults detected by the self-monitoring system of the measuring device are displayed as a diagnostic message in alternation with the operational display.
**==> picture [387 x 119] intentionally omitted <==**
**----- Start of picture text -----**<br>
Diagnostic message<br>Alarm Function check Warning<br>Xxxxxxx Xxxxxxx Xxxxxxx ?<br>?<br>F 1 8 0 C 4 4 6 S 9 6 2<br>T E M P. C I R C . F A I L . I / O 1 O V E R L O A D E M P T Y P I P E<br> A0033011 A0033010 A0033009<br>**----- End of picture text -----**<br>
If two or more diagnostic events are pending simultaneously, only the message of the diagnostic event with the highest priority is shown.
## Diagnostic behavior
|Diagnostic message|Meaning|
|---|---|
||Alarm<br>• The measurement is interrupted.<br>• Signal outputs and totalizers assume the defined alarm condition.<br>• A diagnostic message is generated.|
||Function check<br>Process measured values are simulated to test the outputs/wiring.<br>• IO1/IO2 overload<br>• Flow override is active|
|**?**|Warning<br>• Measurement is resumed.<br>• Measuring operation with limited accuracy<br>• The signal outputs and totalizers are not affected.<br>• A diagnostic message is generated.|
## _Diagnostic behavior of outputs_
|Output|Diagnostic behavior|
|---|---|
|Switch output|• The output is switched off (fail-safe) if events with the status signal F occur<br>• No further response to events with other status signals|
|Pulse output|• Pulse output stops if events with the status signal F occur<br>• No further response to events with other status signals|
|Totalizer|• Totalizer stops if events with the status signal F occur<br>• No further response to events with other status signals|
|Current output|• 3.5 mA is output to signal events with the status signal F<br>• No further response to events with other status signals|
|Voltage output|• 1.75 V is output to signal events with the status signal F<br>• No further response to events with other status signals|
|IO-Link|• All events reported to the Master<br>• Events read and processed further by the Master|
Endress+Hauser
36
Picomag IO-Link
Diagnostics and troubleshooting
## 11.3 Overview of diagnostic events
|Diagnostic<br>event|Event text|Cause|Remedial measures|Status signal [ex-<br>factory]|
|---|---|---|---|---|
|181|Coil. circ. fail.|Coil/frequency failure<br>Coil current PWM<br>outside tolerance range|Replace the measuring<br>device.|F|
|180|Temp. circ.<br>fail.|Temperature sensor<br>open circuit/short-<br>circuit|Replace the measuring<br>device.|F|
|201|Device fail.|Internal hardware error|Replace the measuring<br>device.|F|
|283|Memory fail.|CRC failure|Reset to factory settings.|F|
|446|I/O 1<br>overload|Overload at output 1|Increase load impedance.|C|
|447|I/O 2<br>overload|Overload at output 2|Increase load impedance.|C|
|485|Simulation<br>act.|Measured value<br>simulation active (via<br>remote configuration)|–|C|
|453|Flow override|Flow override active (via<br>auxiliary input)|–|C|
|441|I-Out 1 range|I-output 1 at range limit|Adjust parameter or process.|S|
|444|U-Out 1<br>range|U-output 1 at range<br>limit|Adjust parameter or process.|S|
|443|P-Out 1<br>range|P-output 1 at range<br>limit|Adjust parameter or process.|S|
|442|I-Out 2 range|I-output 2 at range limit|Adjust parameter or process.|S|
|445|U-Out 2<br>range|U-output 2 at range<br>limit|Adjust parameter or process.|S|
|962|Empty pipe|Pipe is completely or<br>partially empty|Adjust the process.|S|
|834|Temperat.<br>range|Medium temperature<br>outside the permitted<br>range|Adjust the process.|S|
For more information on the diagnostics: see the "Diagnostics information" → 51
## 11.4 Device information
The Device info submenu contains all parameters that display different information for device identification.
## Navigation
Menu: "System" → Device info
## _Parameter overview with brief description_
|Parameter|Description|User interface|
|---|---|---|
|Device name|Displays the name of the measuring device.|Picomag|
|Device tag|Shows name of measuring point.|Max. 32 characters, such as letters, numbers or special characters (e.g. @,<br>%, /).|
|Serial number|Displays the serial number of the measuring device.|Max. 11-digit character string comprising letters and numbers.|
Endress+Hauser
37
Diagnostics and troubleshooting
Picomag IO-Link
|Parameter|Description|User interface|
|---|---|---|
|Firmware version|Displays the device firmware version installed.|Character string in the format xx.yy.zz|
|Extended order code|Displays the extended order code.|Character string composed of letters, numbers and certain punctuation<br>marks (e.g. /).|
The following information is shown on the display by tapping once on the housing:
- Status and values for output 1
- Status and values for output 2
- Bluetooth status (On/Off)
- Serial number
- Software version
## 11.5 Firmware history
|Release<br>date|Firmware<br>version|Firmware<br>changes|Documentation type|Documentation|
|---|---|---|---|---|
|09.2017|01.00.zz|Original firmware|Operating<br>Instructions|BA01697D/06/EN/01.17<br>BA01697D/06/EN/02.17<br>BA01697D/06/EN/03.17|
|05.2019|01.01.zz|• Conductivity<br>measurement<br>• Export configuration<br>report<br>• Save/load<br>configuration<br>• Bug fix|Operating<br>Instructions|BA01697D/06/EN/04.19<br>BA01697D/06/EN/05.20<br>BA01697D/06/EN/06.22|
Endress+Hauser
38
Picomag IO-Link
Accessories
## 12 Accessories
Various accessories are available for the device, and can be ordered with the device or at a later stage from Endress+Hauser. An up-to-date overview of accessories is available from your local Endress+Hauser sales organization or on the product page of the Endress +Hauser website: www.endress.com.
_Adapter set (suitable for devices with male thread)_
|Order number|Description|
|---|---|
|71355698|G½" to G³⁄₈" male thread|
|71355699|G½" to R³⁄₈" male thread|
|71355701|G½" to G½" female thread|
|71355702|G½" to R½" male thread|
|71355704|G½" to ½" TriClamp|
|71355705|G³⁄₄" to R³⁄₄" male thread|
|71355707|G³⁄₄" to G³⁄₄" female thread|
|71355708|G³⁄₄" to R³⁄₄" TriClamp|
|71355709|G1" to R1" male thread|
|71355711|G1" to G1" female thread|
|71355712|G1" to 1" TriClamp|
|71355713|G2" to R1½" male thread|
|71355714|G2" to R2" male thread|
|71355717|G2" to G1½" male thread|
|71355718|G2" to G2" female thread|
|71355719|G2" to 2" TriClamp|
|71355720|G2" to 2" Victaulic|
|71399930|G2" to 54mm press fitting|
|71458012|Picomag adapter set G1/2"/R1/2" union nut|
|71458014|Picomag adapter set G3/4"/R1/2" union nut|
## _Cable set_
|Order number|Description|
|---|---|
|71349260|2m/6.5ft, straight, 4x0.34, M12, PUR|
|71349261|5m/16.4ft, straight, 4x0.34, M12, PUR|
|71349262|10m/32.8ft, straight, 4x0.34, M12, PUR|
|71349263|2m/6.5ft, 90 degrees, 4x0.34, M12, PUR|
|71349264|5m/16.4ft, 90 degrees, 4x0.34, M12, PUR|
|71349265|10m/32.8ft, 90 degrees, 4x0.34, M12, PUR|
## _Seal set (suitable for devices with male thread)_
|Order number|Description|
|---|---|
|71354741|DMA15 aramid fiber|
|71354742|DMA20 aramid fiber|
|71354745|DMA25 aramid fiber|
|71354746|DMA50 aramid fiber|
Endress+Hauser
39
Picomag IO-Link
Accessories
## _Ground terminal set_
|Order number|Description|
|---|---|
|71345225|Ground terminal|
Endress+Hauser
40
Picomag IO-Link
Technical data
## 13 Technical data
## 13.1 Input
|13.1<br>Input||
|---|---|
|Measured variables|• Volume flow|
||• Temperature|
||• Conductivity|
|Measuring range|DN 15 (½"): 0.05 to 35 l/min (0.013 to 9.2 gal/min)|
|Volume flow measurement|DN 20 (¾"): 0.1 to 75 l/min (0.026 to 19.8 gal/min)|
||DN 25 (1"): 0.2 to 150 l/min (0.052 to 39.6 gal/min)|
||DN 50 (2"): 1.5 to 750 l/min (0.4 to 198.1 gal/min)|
|Measuring range|–10 to +70 °C (+14 to +158 °F)|
|Medium temperature measurement||
|Measuring range|DN 15 (½"): 20 to 30 000 µS/cm|
|Conductivity measurement|DN 20 (¾"): 20 to 30 000 µS/cm|
||DN 25 (1"): 20 to 30 000 µS/cm|
||DN 50 (2"): 20 to 10 000 µS/cm|
|Digital input|• High or low active|
||• Switch-on level 15 V|
||• Switch-off level 5 V|
||• Internal resistance 7.5 kOhm|
## 13.2 Output
|Output|Max. load|
|---|---|
|Current output|500 Ω|
||Load must not be greater|
|Voltage output|600 Ω|
||Load must not be smaller|
|Pulse output|Max. pulse rate: 10 000 Pulse/s|
|Signal on alarm|• Status signal (as per NAMUR Recommendation NE 107)|
||• Plain text display with remedial action|
|Switch output|• Switching behavior: PNP or NPN|
||• Max. load 250 mA|
## 13.3 Power supply
Supply voltage range 18 to 30 VDC (SELV, PELV, Class 2) Power consumption Maximum 3 W • Without outputs IO1 and IO2: 120 mA • With outputs IO1 and IO2: 120 mA plus the effective load currents
## 13.4 Performance characteristics
|Volume flow measurement||
|---|---|
|Reference operating conditions|Water, +15 to +45 °C, 2 to 6 bar|
|Maximum measured error|±0.8 % o.r. ±0.1 % o.f.s.|
|Repeatability|±0.2 % o.r.|
|Medium temperature measurement||
|Maximum measured error|±2.5 °C|
|Repeatability|±0.5 °C|
|Conductivity measurement||
|Repeatability|±5 %o.r. ±5 µS/cm|
Endress+Hauser
41
Technical data
Picomag IO-Link
Maximum measured error, current output Additional error ±20 µA[ 1)] Repeatability ±10 µA Response time T90 Typically 200 ms 1) At a device temperature of25 °C. Maximum measured error, voltage output Additional error ±60 mV[ 1)] Repeatability ±10 mV Response time T90 Typically 200 ms
- 1) At a device temperature of25 °C.
The response time T90 is the time a measuring system needs to display 90% of the change of the measured value.
## 13.5 Installation
→ 14
## 13.6 Environment
Ambient temperature range –10 to +60 °C (+14 to +140 °F) Storage temperature –25 to +85 °C (–13 to +185 °F) Degree of protection IP65/67, pollution degree 3 Humidity and moisture Suitable for indoor environments with up to 100% rh (wet and damp locations) Operating altitude up to 2 000 m Shock resistance 20 g (11 ms) in accordance with IEC/EN60068-2-27 Vibration resistance Acceleration up to 5 g (10 to 2 000 Hz) in accordance with IEC/ EN60068-2-6 Electromagnetic compatibility (EMC) According to IEC/EN61326 and/or IEC/EN55011 (Class A)
## 13.7 Process
Medium temperature range • –10 to +70 °C (+14 to +158 °F) • Permissible short-term temperature: maximum one hour 85 °C (185 °F) every 4 hours. • Permissible short-term temperature with electronics switched off: maximum one hour 100 °C (212 °F) every 4 hours. Medium properties Liquid, conductivity ≥ 10 µS/cm Pressure Max. 16 barrel
Avoid cavitation in the process.
Magnetism and static electricity
**==> picture [131 x 76] intentionally omitted <==**
_ 8 Avoid magnetic fields_
A0046637
Endress+Hauser
42
Picomag IO-Link
Technical data
## 13.8 Mechanical construction
Measuring device with male thread
**==> picture [368 x 172] intentionally omitted <==**
**----- Start of picture text -----**<br>
A I<br>B<br>K<br>E<br>C D = =<br>d G H<br>F<br>**----- End of picture text -----**<br>
A0033012
## _Dimensions in SI units_
|DN|A|B|C|D|E|||F|G|H|I|K|d|
|---|---|---|---|---|---|---|---|---|---|---|---|---|---|
||[mm]|[mm]|[mm]|[mm]||||[mm]||[mm]|[mm]|[mm]|[mm]|
|15|110|73|40.5|69.5|M12|×|1|43|G½"|56|56|SW 24|12|
|20|110|73|40.5|69.5|M12|×|1|43|G³⁄₄"|56|56|SW 27|15|
|25|110|73|40.5|69.5|M12|×|1|43|G1"|56|56|SW 27|15|
|50|200|113|80|120|M12|×|1|58|G2"|86|86|SW 52|43|
## _Dimensions in US units_
|DN|A|B|C|D|E|||F|G|H|I|K|d|
|---|---|---|---|---|---|---|---|---|---|---|---|---|---|
||[in]|[in]|[in]|[in]||||[in]||[in]|[in]|[in]|[in]|
|15|4.33|2.87|1.59|2.74|M12|×|1|43|G½"|2.2|2.2|AF 1|0.42|
|20|4.33|2.87|1.59|2.74|M12|×|1|43|G³⁄₄"|2.2|2.2|AF 1¹⁄₁₆|0.59|
|25|4.33|2.87|1.59|2.74|M12|×|1|43|G1"|2.2|2.2|AF 1¹⁄₁₆|0.59|
|50|7.87|4.45|3.15|4.72|M12|×|1|58|G2"|3.39|3.39|AF 2¹⁄₁₆|1.69|
## _Weight in SI units_
|DN|[kg]|
|---|---|
|15|0.34|
|20|0.35|
|25|0.36|
|50|1.55|
## _Weight in US units_
|DN|Weight||
|---|---|---|
||[lbs]||
|15|0.75||
|20|0.77||
|25|0.79||
|50|3.42||
|_Materials_|||
|Component||Material|
|Measuring tube||PEEK|
|Electrodes, temperature sensor||1.4435/316L|
|Process connection||1.4404/316L|
Endress+Hauser
43
Technical data
Picomag IO-Link
**==> picture [517 x 262] intentionally omitted <==**
**----- Start of picture text -----**<br>
Component Material<br>Housing 1.4404/316L, 1.4409/CF3M<br>Seal FKM or EPDM<br>Display window Polycarbonate<br>Measuring device with<br>A I<br>female thread<br>B<br>K<br>E<br>C D = =<br>G H<br>F<br>M<br>**----- End of picture text -----**<br>
**==> picture [22 x 4] intentionally omitted <==**
**----- Start of picture text -----**<br>
A0046130<br>**----- End of picture text -----**<br>
## _Dimensions in SI units_
|DN|A|B|C|D|E|||F|G|H|I|K|ØM|
|---|---|---|---|---|---|---|---|---|---|---|---|---|---|
||[mm]|[mm]|[mm]|[mm]||||[mm]||[mm]|[mm]|[mm]|[mm]|
|15|110|73|40.5|69.5|M12|×|1|43|NPT½"|56|56|SW 27|29.5|
|20|110|73|40.5|69.5|M12|×|1|43|NPT³⁄₄"|56|56|SW 32|36|
|25|110|73|40.5|69.5|M12|×|1|43|NPT1"|56|56|SW 41|42|
|50|180|113|80|120|M12|×|1|58|NPT2"|86|86|SW 70|73.5|
_Dimensions in US units_
|DN|A|B|C|D|E|F|G|H|I|K|ØM|
|---|---|---|---|---|---|---|---|---|---|---|---|
||[in]|[in]|[in]|[in]||[in]||[in]|[in]|[in]|[in]|
|15|4.33|2.87|1.59|2.74|M12 × 1|1.69|NPT½"|2.2|2.2|AF 1¹⁄₁₆|1.16|
|20|4.33|2.87|1.59|2.74|M12 × 1|1.69|NPT³⁄₄"|2.2|2.2|AF 1¹⁄₄|1.42|
|25|4.33|2.87|1.59|2.74|M12 × 1|1.69|NPT1"|2.2|2.2|AF 1⁹⁄₁₆|1.65|
|50|7.09|4.45|2.76|4.33|M12 × 1|2.28|NPT2"|3.38|3.38|AF 2³⁄₄|2.89|
|_Weight_|_in SI_|_units_||||||||||
|DN||[kg]||||||||||
|15||0.34||||||||||
|20||0.35||||||||||
|25||0.36||||||||||
|50||1.55||||||||||
_Weight in US units_
|DN|Weight|
|---|---|
||[lbs]|
|15|0.75|
|20|0.77|
|25|0.79|
|50|3.42|
Endress+Hauser
44
Picomag IO-Link
Technical data
_Materials_ Component Material Measuring tube PEEK Electrodes, temperature sensor 1.4435/316L Process connection 1.4404/316L Housing 1.4404/316L, 1.4409/CF3M Seal FKM or EPDM Display window Polycarbonate
## 13.9 Operability
Local display
The device has an onsite display:
**==> picture [130 x 120] intentionally omitted <==**
**----- Start of picture text -----**<br>
4 5 6<br>1 Xxxxxxx<br>2<br>28.56<br>l / min<br>3 23.6<br>°C<br>**----- End of picture text -----**<br>
**==> picture [22 x 4] intentionally omitted <==**
**----- Start of picture text -----**<br>
A0032991<br>**----- End of picture text -----**<br>
- _1 Tag name (configurable)_
- _2 Measured variable 1 (configurable) with sign → 30_
- _3 Measured variable 2 (configurable) with sign → 30_
- _4 Active Bluetooth connection_
- _5 Active I/O-Link connection_
- _6 Device status_
## Display element
3 measured variables can be displayed (volume flow, temperature, conductivity).
Floating point numbers are saved in IEEE754 format and displayed with single precision format. The values are entered in the decimal system, however. As a result, extremely small, insignificant differences can arise between the value entered and the floating point number displayed, depending on the numerical values.
Operation • Via Bluetooth® wireless technology • Via IO-Link Digital communication Via IO-Link SmartBlue app The device has a _Bluetooth_ ® wireless technology interface and can be operated and configured using the SmartBlue app.
- The range under reference conditions is 10 m (33 ft)
- Incorrect operation by unauthorized persons is prevented by means of encrypted communication and password encryption
- The _Bluetooth_ ® wireless technology interface can be disabled
Endress+Hauser
45
Technical data
Picomag IO-Link
## 13.10 Certificates and approvals
Current certificates and approvals for the product are available via the Product Configurator at www.endress.com.
1. Select the product using the filters and search field.
2. Open the product page.
The Configuration button opens the Product Configurator.
CE mark The device meets the legal requirements of the applicable EU Directives. These are listed in the corresponding EU Declaration of Conformity along with the standards applied. Endress+Hauser confirms successful testing of the device by affixing to it the CE mark. UKCA marking The device meets the legal requirements of the applicable UK regulations (Statutory Instruments). These are listed in the UKCA Declaration of Conformity along with the designated standards. By selecting the order option for UKCA marking, Endress+Hauser confirms a successful evaluation and testing of the device by affixing the UKCA mark. Contact address Endress+Hauser UK: Endress+Hauser Ltd. Floats Road Manchester M23 9NF United Kingdom www.uk.endress.com Radio approval The measuring device has radio approval. For detailed information regarding radio approval, see the appendix → 47 Pressure Equipment Devices not bearing this marking (without PED or UKCA) are designed and manufactured Directive according to sound engineering practice. They meet the requirements of a) Art. 4 Para. 3 of the Pressure Equipment Directive 2014/68/EU or b) Part 1, Para. 8 of Statutory Instruments 2016 No. 1105. The scope of application is indicated a) in diagrams 6 to 9 in Annex II of the Pressure Equipment Directive 2014/68/EU or b) Schedule 3, Para. 2 of Statutory Instruments 2016 No. 1105. C[UL] US[ listing] The measuring device is UL-listed. Drinking water approval Availability, see Applicator: https://portal.endress.com/webapp/applicator
Endress+Hauser
46
Picomag IO-Link
Appendix
## 14 Appendix
## 14.1 Radio approvals
## 14.1.1 Argentina
CNC ID: C-22455
## 14.1.2 Brazil
ADENDO AO MANUAL
Modelo: Picomag
**==> picture [70 x 7] intentionally omitted <==**
**----- Start of picture text -----**<br>
04366-18-07311<br>**----- End of picture text -----**<br>
**==> picture [22 x 4] intentionally omitted <==**
**----- Start of picture text -----**<br>
A0037714<br>**----- End of picture text -----**<br>
Para maiores informações, consulte o site da ANATEL: www.gov.br/anatel/pt-br
Este equipamento não tem direito à proteção contra interferência prejudicial e não pode causar interferência em sistemas devidamente autorizados.
## 14.1.3 Europe
This device meets the requirements of the Telecommunications Directive RED 2014/53/EU.
## 14.1.4 India
ETA Certificate No: ETA - 1707/18-RLO(NE)
Endress+Hauser
47
Appendix
Picomag IO-Link
## 14.1.5 Indonesia
**==> picture [172 x 159] intentionally omitted <==**
**----- Start of picture text -----**<br>
76485/SDPPI/2021<br>4962<br>Dilarang melakukan<br>perubahan spesifikasi<br>yang dapat<br>menimbulkan<br>gangguan fisik dan/<br>atau elektromagnetik<br>terhadap lingkungan<br>sekitarnya.<br>**----- End of picture text -----**<br>
## 14.1.6 Canada and United States
## English
This device complies with Part 15 of the FCC Rules and with Industry Canada licenceexempt RSS standard(s).
Operation is subject to the following two conditions:
- This device may not cause harmful interference, and
- This device must accept any interference received, including interference that may cause undesired operation.
- Changes or modifications made to this equipment not expressly approved by Endress+Hauser Flowtec AG may void the user's authorization to operate this equipment.
## Français
Le présent appareil est conforme aux CNR d'industrie Canada applicables aux appareils radio exempts de licence.
L'exploitation est autorisée aux deux conditions suivantes :
- L'appareil ne doit pas produire de brouillage, et
- L'utilisateur de l'appareil doit accepter tout brouillage radioélectrique subi, même si le brouillage est susceptible d'en compromettre le fonctionnement.
- Les changements ou modifications apportés à cet appareil non expressément approuvés par Endress+Hauser Flowtec AG peuvent annuler l'autorisation de l'utilisateur d'opérer cet appareil.
## 14.1.7 Mexico
La operación de este equipo está sujeta dos condiciones: (1) es posible que este equipo o dispositivo no cause interferencia perjudicial y (2) este equipo o dispositivo debe aceptar cualquier interferencia, incluyendo la que pueda causar su operación no deseada.
**==> picture [387 x 61] intentionally omitted <==**
**----- Start of picture text -----**<br>
IFT No.: RCPENPI21-3683<br> A0034100<br>**----- End of picture text -----**<br>
Endress+Hauser
48
Picomag IO-Link
Appendix
## 14.1.8 Republic of Korea
_KC_ 인증
적합성평가정보
R-C-EH7-Picomag
상호 : 한국엔드레스하우저 주식회사
기자재명칭(모델명): 특정소출력 무선기기(무선데이터통신시스템용무선기기) / Picomag 제조국 및 제조국가 : Endress+Hauser Flowtec AG / 프랑스 제조년월 : 제조년월로 표기
*사용자안내문
이 기기는 업무용 환경에서 사용할 목적으로 적합성평가를 받은 기기로서 가정용 환 경에서 사용하는 경우 전파간섭의 우려가 있습니다.
## 14.1.9 Singapore
Complies with IMDA Standards DA108204
## 14.1.10 Taiwan
## 低功率電波輻射性電機管理辦法
第十二條 經型式認證合格之低功率射頻電機,非經許可,公司、商號或使用者均不得擅自變更頻 率、加大功率或變更原設計之特性及功能。 第十四條 低功率射頻電機之使用不得影響飛航安全及干擾合法通信;經發現有干擾現象時,應改善 至無干擾時方得繼續使用。前項合法通信,指依電信法規定作業之無線電通信。低功率射 頻電機須忍受合法通信或工業、科學及醫療用電波輻射性電機設備之干擾。 產品名稱 Endress + Hauser 產品型號 Picomag 產地 瑞士 製造商 Endress + Hauser Flowtec AG
## 14.1.11 Thailand
เครื่องโทรคมนาคมและอุปกรณนี้ มีความสอดคลองตามขอกำหนดของ กสทช.
(This telecommunication equipment is in compliance with NBTC requirements.)
Endress+Hauser
49
Appendix
Picomag IO-Link
**==> picture [22 x 4] intentionally omitted <==**
**----- Start of picture text -----**<br>
A0041612<br>**----- End of picture text -----**<br>
## 14.1.12 United Arab Emirates
TRA Registered ER68711/19
## 14.1.13 Other countries
Other national approvals are available on request.
Endress+Hauser
50
Picomag IO-Link
Appendix
## 14.2 IO-Link process data
## 14.2.1 Data structure
|Bit<br>number|119…<br>112|111…<br>104|103…<br>96|95…<br>88|87…<br>80|79…<br>72|71…<br>64|63…<br>56|55…<br>48|47…<br>40|39…<br>32|31…<br>24|23…<br>16|15…<br>8|7…0|
|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|
|Data|Conductivity in µS/cm||||Totalizer in l||||Volume flow in l/s||||Temperature<br>in ¹⁄₁₀ °C||Status|
|Data<br>type|32-bit floating point<br>number with single<br>precision (IEEE 754)||||32-bit floating point<br>number with single<br>precision (IEEE 754)||||32-bit floating point<br>number with single<br>precision (IEEE 754)||||16-bit two's<br>complement||8-bit|
_Data structure of the status bits 7 to 0_
|Bit|Description|
|---|---|
|0|Switches once per sampling rate|
|1|Reserved|
|2|Current status S-Out 1|
|3|Current status S-Out 2|
|4|Reserved|
|5|Reserved|
|6|Reserved|
|7|Reserved|
## 14.2.2 Diagnostic information
|Diagnostic code|Diagnostic code|Display text|Coding<br>|PDValid<br>|Priority|
|---|---|---|---|---|---|
|Status<br>NE 107|Diagnostic number||(hex)|Validity||
|–||SYSTEM OK|0x0000|1|1|
|F|181|COIL CIRC.FAIL.|0x5000|0|2|
|F|180|TEMP.CIRC.FAIL.|0x5000|0|3|
|F|201|DECICE FAIL.|0x5000|0|4|
|F|283|MEMORY FAIL.|0x8C00|0|5|
|C|446|I/O 1 OVERLOAD|0x180C|1|6|
|C|447|I/O 2 OVERLOAD|0x180C|1|7|
|C|485|SIMULATION ACT.|0x8C01|1|8|
|C|453|FLOW OVERRIDE|0x180D|1|9|
|S|441|I-OUT 1 RANGE|0x180A|1|10|
|S|444|U-OUT 1 RANGE|0x1809|1|11|
|S|443|P-OUT 1 RANGE|0x180B|1|12|
|S|442|I-OUT 2 RANGE|0x180A|1|13|
|S|445|U-OUT 2 RANGE|0x1809|1|14|
|S|962|EMPTY PIPE|0x180E|1|15|
|S|834|TEMPERAT. RANGE|0x8C20|1|16|
Endress+Hauser
51
Appendix
Picomag IO-Link
## 14.3 IO-Link ISDU parameter list
The individual parts of a parameter description are described in the following section:
|Designation|ISDU<br>(hex)|ISDU<br>(dec)|Size<br>(byte)|Data type|Access|Value range|Factory setting|Range<br>limits|
|---|---|---|---|---|---|---|---|---|
|Identification|||||||||
|Device Tag<br>First 10 characters displayed<br>(starting from left)|0x0018|24|32 (max.)|string|r/w||EH_DMA_XXZZ||
|Device Name|0x0012|18|16 (max.)|string|r||Picomag||
|Device ID1|0x0009|9|1|uint|r||0x01||
|Device ID2|0x000A|10|1|uint|r||0x01||
|Device ID3|0x000B|11|1|uint|r||0x00||
|Vendor Name|0x0010|16|32 (max.)|string|r||Endress+Hauser||
|Vendor ID1|0x0007|7|1|uint|r||0x00||
|Vendor ID2|0x0008|8|1|uint|r||0x11||
|Device Serial No.<br>e.g. (YMXXXXZZ)|0x0015|21|11 (max.)|string|r||see nameplate||
|Firmware Version<br>e.g. 01.00.00|0x0017|23|8 (max.)|string|r||||
|Order Code<br>e.g. DMA15-AAAAA1|0x0102|258|18 (max.)|string|r||see nameplate||
|Device Type|0x0100|256|2|uint|r||0x94FF||
|Diagnostics|||||||||
|Actual Diagnostics<br>e.g. C485 (= SIMULATION ACT.)|0x0104|260|4|string|r||||
|Last Diagnostics<br>e.g. S962 (= EMPTY PIPE)|0x0105|261|4|string|r||||
|Simulation Proc. Var.|0x015F|351|2|uint|r/w|enable=1<br>disable=0|||
|Sim.Proc.Var.Value Volumeflow<br>Unit selection list fromUnit<br>Volumeflow|0x0166|358|4|float|r/w||0.0|-10⁶<br>10⁶|
|Sim.Proc.Var.Value<br>Temperature<br>Unit selection list fromUnit<br>Temperature|0x0168|360|4|float|r/w||0.0|-10⁴<br>10⁴|
|Sim.Proc.Var.Value Conductivity<br>Unit selection list fromUnit<br>Conductivity|0x0167|359|4|float|r/w||0.0|0<br>10⁶|
|Measured Values|||||||||
|Volumeflow<br>Current volume flow measured<br>value|0x0161|353|4|float|r||||
|Temperature<br>Current temperature measured<br>value|0x0163|355|4|float|r||||
|Conductivity<br>Current conductivity measured<br>value|0x0164|365|4|float|r||||
|Totalizer<br>Current totalizer measured value|0x0169|361|4|float|r/w||0.0||
Endress+Hauser
52
Picomag IO-Link
Appendix
|Designation|ISDU<br>(hex)|ISDU<br>(dec)|Size<br>(byte)|Data type|Access|Value range|Factory setting|Range<br>limits|
|---|---|---|---|---|---|---|---|---|
|Totalizer Reset|0x016A|362|2|uint|w|cancel=0<br>reset=1|cancel||
|System Units|||||||||
|Unit Volumeflow|0x0226|550|2|uint|r/w|l/s=0<br>l/h=5<br>fl. oz/min=4<br>m³/h=1<br>l/min=2<br>Usgpm=3|l/min||
|Unit Volume|0x0227|551|2|uint|r/w|ml=0<br>USozf=1<br>l=2<br>m³=3<br>Usgal=4|ml||
|Unit Temperature|0x0228|552|2|uint|r/w|°C=0<br>°F=1|°C||
|Unit Conductivity|0x0229|553|2|uint|r/w|µS/cm=0<br>S/m=1<br>mS/cm=2|µS/cm=0||
|Unit Totalizer|0x016B|363|2|uint|r/w|USozf=1<br>l=2<br>m³=3<br>Usgal=4<br>kl=5<br>Ml=6<br>kUsg=7|m³||
|Sensor|||||||||
|Install. Direction<br>In relation to direction of arrow on<br>the device|0x015E|350|2|uint|r/w|forward=0<br>reverse=1|forward||
|Low Flow Cut Off<br>The flow rate below the selected<br>value is zero<br>Unit selection list fromUnit<br>Volumeflow|0x0160|352|4|float|r/w||0.4/0.75/1.2/5.0<br>l/min|0<br>10⁶|
|Damping<br>Volume flow damping via the PT1<br>element<br>Unit: s|0x01A4|420|4|float|r/w||0 s|0<br>100|
Endress+Hauser
53
Appendix
Picomag IO-Link
|Designation|ISDU<br>(hex)|ISDU<br>(dec)|Size<br>(byte)|Data type|Access|Value range|Factory setting|Range<br>limits|
|---|---|---|---|---|---|---|---|---|
|Output 1|||||||||
|Operating Mode<br>IO-Link is set if connected to a<br>master|0x01F4|500|2|uint|r/w|P-Out=0<br>I-Out=1<br>S-In=2<br>S-Out=3<br>IO-Link=4<br>U-Out=5<br>off=6|IO-Link||
|Current output I-Out 1|||||||||
|I - OUT Assign|0x0258|600|2|uint|r/w|off=0<br>volume flow=1<br>temperature=2<br>conductivity=4|volume flow||
|Q-Start-Value<br>ASP1)for volume flow<br>Unit selection list fromUnit<br>Volumeflow|0x0259|601|4|float|r/w||0 l/min|-9.9·10⁹<br>9.9·10⁹|
|Q-End-Value<br>AEP2)for volume flow<br>Unit selection list fromUnit<br>Volumeflow|0x025A|602|4|float|r/w||25/50/100/750<br>l/min|-9.9·10⁹<br>9.9·10⁹|
|T-Start-Value<br>ASP1)for temperature<br>Unit selection list fromUnit<br>Temperature|0x025F|607|4|float|r/w||-10 °C|-9.9·10⁹<br>9.9·10⁹|
|T-End-Value<br>AEP2)for temperature<br>Unit selection list fromUnit<br>Temperature|0x0260|608|4|float|r/w||+70 °C|-9.9·10⁹<br>9.9·10⁹|
|s-Start-Value<br>ASP1)for conductivity<br>Unit selection list fromUnit<br>Conductivity|0x025D|605|4|float|r/w||0|-9.9·10⁹<br>9.9·10⁹|
|s-End-Value<br>AEP2)for conductivity<br>Unit selection list fromUnit<br>Conductivity|0x025E|606|4|float|r/w||1000|-9.9·10⁹<br>9.9·10⁹|
Endress+Hauser
54
Picomag IO-Link
Appendix
|Designation|ISDU<br>(hex)|ISDU<br>(dec)|Size<br>(byte)|Data type|Access|Value range|Factory setting|Range<br>limits|
|---|---|---|---|---|---|---|---|---|
|Pulse output P-Out|||||||||
|Pulse Value<br>Unit selection list fromUnit<br>Volume|0x03E8|1000|4|float|r/w||0.5/1.0/2.0/10.0 ml|10¯⁹<br>9.9·10⁹|
|Switch output S-Out 1|||||||||
|Switch Polarity|0x032B|811|2|uint|r/w|PNP=0<br>NPN=1|PNP||
|Switch Function|0x0320|800|2|uint|r/w|alarm=0<br>off=1<br>on=2<br>lim.vol.flow=3,<br>lim.temp.=4<br>lim.vol.=5<br>lims=11<br>win.vol.flow=6<br>win.temp.=7<br>win.vol.=8<br>wins=13<br>epd=9|alarm||
|Q-ON-Value<br>Unit selection list fromUnit<br>Volumeflow|0x0321|801|4|float|r/w||20/40/80/600 l/min|-9.9·10⁹<br>9.9·10⁹|
|Q-OFF-Value<br>Unit selection list fromUnit<br>Volumeflow|0x0322|802|4|float|r/w||15/30/60/450 l/min|-9.9·10⁹<br>9.9·10⁹|
|T-ON-Value<br>Unit selection list fromUnit<br>Temperature|0x0327|807|4|float|r/w||+ 60 °C|-9.9·10⁹<br>9.9·10⁹|
|T-OFF-Value<br>Unit selection list fromUnit<br>Temperature|0x0328|808|4|float|r/w||+ 50 °C|-9.9·10⁹<br>9.9·10⁹|
|V-ON-Value<br>Unit selection list fromUnit<br>Totalizer|0x0329|809|4|float|r/w||0.2/0.4/0.8/6.0 m³|-9.9·10⁹<br>9.9·10⁹|
|V-OFF-Value<br>Unit selection list fromUnit<br>Totalizer|0x032A|810|4|float|r/w||0.15/0.3/0.6/4.5 m³|-9.9·10⁹<br>9.9·10⁹|
|s-ON-Value<br>Unit selection list fromUnit<br>Conductivity|0x0325|805|4|float|r/w||500|-9.9·10⁹<br>9.9·10⁹|
|s-OFF-Value<br>Unit selection list fromUnit<br>Conductivity|0x0326|806|4|float|r/w||200|-9.9·10⁹<br>9.9·10⁹|
|Voltage output U-Out 1|||||||||
|U - OUT Assign|0x02BC|700|2|uint|r/w|off=0<br>volume flow=1<br>temperature=2<br>conductivity=4|volume flow||
|Q-Start-Value<br>ASP1)for volume flow<br>Unit selection list fromUnit<br>Volumeflow|0x02BD|701|4|float|r/w||0 l/min||
|Q-End-Value<br>AEP2)for volume flow<br>Unit selection list fromUnit<br>Volumeflow|0x02BE|702|4|float|r/w||25/50/100/750<br>l/min||
Endress+Hauser
55
Appendix
Picomag IO-Link
|Designation|ISDU<br>(hex)|ISDU<br>(dec)|Size<br>(byte)|Data type|Access|Value range|Factory setting|Range<br>limits|
|---|---|---|---|---|---|---|---|---|
|T-Start-Value<br>ASP1)for temperature<br>Unit selection list fromUnit<br>Temperature|0x02C3|707|4|float|r/w||-10 °C||
|T-End-Value<br>AEP2)for temperature<br>Unit selection list fromUnit<br>Temperature|0x02C4|708|4|float|r/w||+70 °C||
|s-Start-Value<br>ASP1)for conductivity<br>Unit selection list fromUnit<br>Conductivity|0x02C1|705|4|float|r/w||0 µS/cm||
|s-End-Value<br>AEP2)for conductivity<br>Unit selection list fromUnit<br>Conductivity|0x02C2|706|4|float|r/w||1000 µS/cm||
|Digital input D-In 1|||||||||
|D-IN Polarity|0x0385|901|2|uint|r/w|low=0<br>high=1|high||
|D-IN Function|0x0384|900|2|uint|r/w|off=0<br>res.tot.=1<br>zero ret.=2|res.tot.||
|IO-Link|||||||||
|IO-LINK Vendor Name|0x0010|16|32 (max.)|string|r||Endress+Hauser||
|IO-LINK Product Name|0x0012|18|16 (max.)|string|r||Picomag||
|IO-LINK RevisionID|0x0004|4|1|uint|r||0x11||
|Output 2|||||||||
|Operating Mode|0x01F5|501|2|uint|r/w|I-Out=1<br>S-In=2<br>S-Out=3<br>U-Out=5<br>off=6|off||
|Current output I-Out 2|||||||||
|I - OUT Assign|0x028A|650|2|uint|r/w|off=0<br>volume flow=1<br>temperature=2|temperature||
|Q-Start-Value<br>ASP1)for volume flow<br>Unit selection list fromUnit<br>Volumeflow|0x028B|651|4|float|r/w||0 l/min||
|Q-End-Value<br>AEP2)for volume flow<br>Unit selection list fromUnit<br>Volumeflow|0x028C|652|4|float|r/w||25/50/100/750<br>l/min||
|T-Start-Value<br>ASP1)for temperature<br>Unit selection list fromUnit<br>Temperature|0x0291|657|4|float|r/w||-10 °C||
|T-End-Value<br>AEP2)for temperature<br>Unit selection list fromUnit<br>Temperature|0x0292|658|4|float|r/w||+70 °C||
|s-Start-Value<br>ASP1)for conductivity<br>Unit selection list fromUnit<br>Conductivity|0x028F|655|4|float|r/w||0|-9.9·10⁹<br>9.9·10⁹|
Endress+Hauser
56
Picomag IO-Link
Appendix
|Designation|ISDU<br>(hex)|ISDU<br>(dec)|Size<br>(byte)|Data type|Access|Value range|Factory setting|Range<br>limits|
|---|---|---|---|---|---|---|---|---|
|s-End-Value<br>AEP2)for conductivity<br>Unit selection list fromUnit<br>Conductivity|0x0290|656|4|float|r/w||1000|-9.9·10⁹<br>9.9·10⁹|
|Switch output S-Out 2|||||||||
|Switch Polarity|0x035D|861|2|uint|r/w|PNP=0<br>NPN=1|PNP||
|Switch Function|0x0352|850|2|uint|r/w|alarm=0<br>off=1<br>on=2<br>lim.vol.flow=3<br>lim.temp.=4<br>lim.vol.=5,<br>win.vol.flow=6<br>lims=11<br>win.temp.=7<br>win.vol.=8<br>wins=13<br>epd=9|alarm||
|Q-ON-Value<br>Unit selection list fromUnit<br>Volumeflow|0x0353|851|4|float|r/w||20/40/80/600 l/min||
|Q-OFF-Value<br>Unit selection list fromUnit<br>Volumeflow|0x0354|852|4|float|r/w||15/30/60/450 l/min||
|T-ON-Value<br>Unit selection list fromUnit<br>Temperature|0x0359|857|4|float|r/w||+ 60 °C||
|T-OFF-Value<br>Unit selection list fromUnit<br>Temperature|0x035A|858|4|float|r/w||+ 50 °C||
|V-ON-Value<br>Unit selection list fromTotalizer|0x035B|859|4|float|r/w||0.2/0.4/0.8/6.0 m³||
|V-OFF-Value<br>Unit selection list fromTotalizer|0x035C|860|4|float|r/w||0.15/0.3/0.6/4.5 m³||
|s-ON-Value<br>Unit selection list from<br>Conductivity|0x0357|855|4|float|r/w||500||
|s-OFF-Value<br>Unit selection list from<br>Conductivity|0x0358|856|4|float|r/w||200||
|Voltage output U-Out 2|||||||||
|U - OUT Assign|0x02EE|750|2|uint|r/w|off=0<br>volume flow=1<br>temperature=2|temperature||
|Q-Start-Value<br>ASP1)for volume flow<br>Unit selection list fromUnit<br>Volumeflow|0x02EF|751|4|float|r/w||0 l/min||
|Q-End-Value<br>AEP2)for volume flow<br>Unit selection list fromUnit<br>Volumeflow|0x02F0|752|4|float|r/w||25/50/100/750<br>l/min||
|T-Start-Value<br>ASP1)for temperature fromUnit<br>Temperature|0x02F5|757|4|float|r/w||-10 °C||
Endress+Hauser
57
Appendix
Picomag IO-Link
|Designation|ISDU<br>(hex)|ISDU<br>(dec)|Size<br>(byte)|Data type|Access|Value range|Factory setting|Range<br>limits|
|---|---|---|---|---|---|---|---|---|
|T-End-Value<br>AEP2)for temperature fromUnit<br>Temperature|0x02F6|758|4|float|r/w||+70 °C||
|s-Start-Value<br>ASP1)for temperature from<br>Conductivity|0x02F3|755|4|float|r/w||0 µS/cm||
|s-End-Value<br>AEP2)for temperature from<br>Conductivity|0x02F4|756|4|float|r/w||1000 µS/cm||
|Digital input D-In 2|||||||||
|D-IN Polarity|0x0395|917|2|uint|r/w|low=0<br>high=1|high||
|D-IN Function|0x0394|916|2|uint|r/w|off=0<br>res.tot.=1<br>zero ret.=2|res.tot.||
|Display|||||||||
|Display Layout|0x01C3|451|2|uint|r/w|QV=0<br>QT=1<br>Qs=3<br>VT=2, Vs=4<br>Ts=5<br>QVTs=6<br>QVTs_m=7|QT||
|Display Rotation|0x01C4|452|2|uint|r/w|0 °=0<br>90 °=1<br>180 °=2<br>270 °=3<br>auto=4|Auto||
|Display Backlight|0x01C2|450|2|uint|r/w|0…100|50||
|Bluetooth configuration|||||||||
|Bluetooth Function|0x041A|1050|2|uint|r/w|on=1<br>off=0|on||
|Bluetooth Tx Pwr Level|0x041B|1051|2|uint|r|0…4|||
|Bluetooth Conn. Status|0x041C|1052|1|uint|r||||
|Administration|||||||||
|Set Access Code<br>Define the access code|0x0108|264|2|uint|w||0000||
|Access Code<br>Enter the access code|0x0107|263|2|uint|w||||
|Reset Device|0x010E|270|2|uint|w|cancel=0<br>restore fact.=1<br>restart=4|cancel||
|Product Specific Process Values|||||||||
|Status IO 1|0x0386|902|2|uint|r|low=0<br>high=1|||
|Status IO 2|0x0396|918|2|uint|r|low=0<br>high=1|||
- 1) Analog Start Point
- 2) Analog End Point
Endress+Hauser
58
Picomag IO-Link
Index
## Index
## A
**==> picture [248 x 305] intentionally omitted <==**
**----- Start of picture text -----**<br>
||||||||||||||||||||||||||||||||||||
|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|
|About this document .|.|.|.|.|.|.|.|.|.|.|.|.|.|.|.|.|.|.|.|.|.|.|.|.|.|5|
|Accessories .|.|.|.|.|.|.|.|.|.|.|.|.|.|.|.|.|.|.|.|.|.|.|.|.|.|.|.|.|.|.|.|39|
|Appendix .|.|.|.|.|.|.|.|.|.|.|.|.|.|.|.|.|.|.|.|.|.|.|.|.|.|.|.|.|.|.|.|.|47|
|Application|
|Media .|.|.|.|.|.|.|.|.|.|.|.|.|.|.|.|.|.|.|.|.|.|.|.|.|.|.|.|.|.|.|.|.|.|8|
|B|
|Basic safety instructions .|.|.|.|.|.|.|.|.|.|.|.|.|.|.|.|.|.|.|.|.|.|.|8|
|Bidirectional flow measurement (Q), temperature|
|measurement (T) .|.|.|.|.|.|.|.|.|.|.|.|.|.|.|.|.|.|.|.|.|.|.|.|. 25, 28|
|C|
|CE mark .|.|.|.|.|.|.|.|.|.|.|.|.|.|.|.|.|.|.|.|.|.|.|.|.|.|.|.|.|.|.|.|9, 46|
|Commissioning .|.|.|.|.|.|.|.|.|.|.|.|.|.|.|.|.|.|.|.|.|.|.|.|.|.|.|.|.|22|
|Configuring system units .|.|.|.|.|.|.|.|.|.|.|.|.|.|.|.|.|.|.|.|.|.|22|
|Configuring the current output .|.|.|.|.|.|.|.|.|.|.|.|.|.|.|.|.|24|
|Configuring the digital input .|.|.|.|.|.|.|.|.|.|.|.|.|.|.|.|.|.|.|28|
|Configuring the display .|.|.|.|.|.|.|.|.|.|.|.|.|.|.|.|.|.|.|.|.|.|.|30|
|Configuring the IO modules .|.|.|.|.|.|.|.|.|.|.|.|.|.|.|.|.|.|.|.|24|
|Configuring the measuring device .|.|.|.|.|.|.|.|.|.|.|.|.|.|.|22|
|Configuring the pulse output .|.|.|.|.|.|.|.|.|.|.|.|.|.|.|.|.|.|.|25|
|Configuring the voltage output .|.|.|.|.|.|.|.|.|.|.|.|.|.|.|.|.|27|
|Connecting requirements .|.|.|.|.|.|.|.|.|.|.|.|.|.|.|.|.|.|.|.|.|.|16|
|Connecting the measuring device .|.|.|.|.|.|.|.|.|.|.|.|.|.|.|.|19|
|Current output configuration version .|.|.|.|.|.|.|.|.|.|.|.|.|18|
|Current version data for the device .|.|.|.|.|.|.|.|.|.|.|.|.|.|.|21|
**----- End of picture text -----**<br>
## D
Data management . . . . . . . . . . . . . . . . . . . . . . . . . . . 31 Declaration of Conformity . . . . . . . . . . . . . . . . . . . . . . 9 Device information . . . . . . . . . . . . . . . . . . . . . . . . . . 37 Device master file . . . . . . . . . . . . . . . . . . . . . . . . . . . 21 Diagnostic behavior . . . . . . . . . . . . . . . . . . . . . . . . . . 36 Diagnostic information on local display . . . . . . . . . . . 36 Diagnostic message . . . . . . . . . . . . . . . . . . . . . . . . . . 36 Diagnostics and troubleshooting . . . . . . . . . . . . . . . . 35 Digital input configuration version . . . . . . . . . . . . . . . 18 Dimensions in SI units . . . . . . . . . . . . . . . . . . . . . 43, 44 Dimensions in US units . . . . . . . . . . . . . . . . . . . . . . . 43 Document function . . . . . . . . . . . . . . . . . . . . . . . . . . . 5 Documentation . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6 Drinking water approval . . . . . . . . . . . . . . . . . . . . . . 46 E Electrical connection . . . . . . . . . . . . . . . . . . . . . . . . . 16
## F
Field of application Residual risks . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 8 Firmware history . . . . . . . . . . . . . . . . . . . . . . . . . . . . 38
## I
**==> picture [248 x 44] intentionally omitted <==**
**----- Start of picture text -----**<br>
||||||||||||||||||||||||||||||||
|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|
|Identification .|.|.|.|.|.|.|.|.|.|.|.|.|.|.|.|.|.|.|.|.|.|.|.|.|.|.|.|.|.|22|
|Identifying the measuring device .|.|.|.|.|.|.|.|.|.|.|.|.|.|.|.|11|
|Incoming acceptance .|.|.|.|.|.|.|.|.|.|.|.|.|.|.|.|.|.|.|.|.|.|.|.|.|11|
|Inlet and outlet runs .|.|.|.|.|.|.|.|.|.|.|.|.|.|.|.|.|.|.|.|.|.|.|.|.|14|
**----- End of picture text -----**<br>
## Inspection
**==> picture [247 x 141] intentionally omitted <==**
**----- Start of picture text -----**<br>
|||||||||||||||||||||||||||||||||||
|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|
|Received goods .|.|.|.|.|.|.|.|.|.|.|.|.|.|.|.|.|.|.|.|.|.|.|.|.|.|.|11|
|Intended use .|.|.|.|.|.|.|.|.|.|.|.|.|.|.|.|.|.|.|.|.|.|.|.|.|.|.|.|.|.|.|.|8|
|IO-Link configuration version .|.|.|.|.|.|.|.|.|.|.|.|.|.|.|.|.|.|19|
|M|
|Magnetism .|.|.|.|.|.|.|.|.|.|.|.|.|.|.|.|.|.|.|.|.|.|.|.|.|.|.|.|.|.|.|.|42|
|Materials .|.|.|.|.|.|.|.|.|.|.|.|.|.|.|.|.|.|.|.|.|.|.|.|.|.|.|.|.|.|43, 45|
|Mounting .|.|.|.|.|.|.|.|.|.|.|.|.|.|.|.|.|.|.|.|.|.|.|.|.|.|.|.|.|.|.|.|.|14|
|Mounting location .|.|.|.|.|.|.|.|.|.|.|.|.|.|.|.|.|.|.|.|.|.|.|.|.|.|.|14|
|Mounting position .|.|.|.|.|.|.|.|.|.|.|.|.|.|.|.|.|.|.|.|.|.|.|.|.|.|.|14|
|Mounting requirements .|.|.|.|.|.|.|.|.|.|.|.|.|.|.|.|.|.|.|.|.|.|.|14|
|Mounting the measuring device .|.|.|.|.|.|.|.|.|.|.|.|.|.|.|.|.|14|
**----- End of picture text -----**<br>
## O
Operation . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 33 Operation options . . . . . . . . . . . . . . . . . . . . . . . . . . . 20 SmartBlue app . . . . . . . . . . . . . . . . . . . . . . . . . . . 20 Operational safety . . . . . . . . . . . . . . . . . . . . . . . . . . . . 9 Overview of device description files . . . . . . . . . . . . . . 21 Overview of diagnostic events . . . . . . . . . . . . . . . . . . 37 Overview of the operating menu . . . . . . . . . . . . . . . . 22
## P
**==> picture [247 x 59] intentionally omitted <==**
**----- Start of picture text -----**<br>
|||||||||||||||||||||||||||||||||
|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|
|Packaging disposal .|.|.|.|.|.|.|.|.|.|.|.|.|.|.|.|.|.|.|.|.|.|.|.|.|.|13|
|Pin assignment, device plug .|.|.|.|.|.|.|.|.|.|.|.|.|.|.|.|.|.|.|.|16|
|Post-connection check .|.|.|.|.|.|.|.|.|.|.|.|.|.|.|.|.|.|.|.|.|.|.|.|19|
|Pressure Equipment Directive .|.|.|.|.|.|.|.|.|.|.|.|.|.|.|.|.|.|46|
|Product safety .|.|.|.|.|.|.|.|.|.|.|.|.|.|.|.|.|.|.|.|.|.|.|.|.|.|.|.|.|.|.|9|
**----- End of picture text -----**<br>
## R
Radio approval . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 46 Radio approvals . . . . . . . . . . . . . . . . . . . . . . . . . . 47 Registered trademarks . . . . . . . . . . . . . . . . . . . . . . . . . 6 Requirements for connecting cable . . . . . . . . . . . . . . . 16 Requirements for personnel . . . . . . . . . . . . . . . . . . . . . 8 S Security . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 31 Setting the installation direction and measurement . . 23 Static electricity . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 42 Storage conditions . . . . . . . . . . . . . . . . . . . . . . . . . . . 13 Storage temperature . . . . . . . . . . . . . . . . . . . . . . . . . 13 Switch output configuration version . . . . . . . . . . . . . . 17 Switch/pulse output configuration version . . . . . . . . . 17 Switching ON the measuring device . . . . . . . . . . . . . . 22 Symbols used . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5 System integration . . . . . . . . . . . . . . . . . . . . . . . . . . 21
## T
**==> picture [247 x 82] intentionally omitted <==**
**----- Start of picture text -----**<br>
||||||||||||||||||||||||||||||||||||
|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|
|Technical data .|.|.|.|.|.|.|.|.|.|.|.|.|.|.|.|.|.|.|.|.|.|.|.|.|.|.|.|.|.|41|
|Certificates and approvals .|.|.|.|.|.|.|.|.|.|.|.|.|.|.|.|.|.|.|46|
|Environment .|.|.|.|.|.|.|.|.|.|.|.|.|.|.|.|.|.|.|.|.|.|.|.|.|.|.|.|42|
|Input .|.|.|.|.|.|.|.|.|.|.|.|.|.|.|.|.|.|.|.|.|.|.|.|.|.|.|.|.|.|.|.|.|.|41|
|Installation .|.|.|.|.|.|.|.|.|.|.|.|.|.|.|.|.|.|.|.|.|.|.|.|.|.|.|.|.|42|
|Mechanical construction .|.|.|.|.|.|.|.|.|.|.|.|.|.|.|.|.|43, 44|
|Operability .|.|.|.|.|.|.|.|.|.|.|.|.|.|.|.|.|.|.|.|.|.|.|.|.|.|.|.|.|.|45|
**----- End of picture text -----**<br>
Endress+Hauser
59
Index
Picomag IO-Link
|Output . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 41|Output . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 41|
|---|---|
|Performance characteristics . . . . . . . . . . . . . . .|. . 41|
|Power supply . . . . . . . . . . . . . . . . . . . . . . . . . . . . 41||
|Process . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .|. . 42|
|Temperature range||
|Storage temperature . . . . . . . . . . . . . . . . . . . .|. . . 13|
|Tools||
|Transport . . . . . . . . . . . . . . . . . . . . . . . . . . . .|. . . 13|
|Totalizer . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 29||
|Transporting the measuring device . . . . . . . . . . . .|. . 13|
|Troubleshooting||
|General . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .|. . 35|
|U||
|UKCA marking . . . . . . . . . . . . . . . . . . . . . . . . . . .|. . 46|
|Unidirectional flow measurement (Q), conductivity||
|measurement . . . . . . . . . . . . . . . . . . . . . . . . . . .|25, 28|
|Using the measuring device||
|Borderline cases . . . . . . . . . . . . . . . . . . . . . . . . . . . 8||
|Incorrect use . . . . . . . . . . . . . . . . . . . . . . . . . .|. . . 8|
|V||
|Voltage output configuration version . . . . . . . . . . . . . 18||
|W||
|W@M Device Viewer . . . . . . . . . . . . . . . . . . . . . .|. . 11|
|Weight||
|Transport (notes) . . . . . . . . . . . . . . . . . . . . . . . . . 13||
|Weight in SI units . . . . . . . . . . . . . . . . . . . . . . . .|43, 44|
|Weight in US units . . . . . . . . . . . . . . . . . . . . . . .|43, 44|
|Workplace safety . . . . . . . . . . . . . . . . . . . . . . . . .|. . .<br>8|
Endress+Hauser
60
*71552010* 71552010
www.addresses.endress.com
Updated at June 9, 2026
About Novapart
Novapart is a B2B electronic component broker specialising in stock shortages and cost reduction. We source hard-to-find parts and identify compliant alternatives across a catalogue of 410,000+ components from 500+ manufacturers.
Learn more →Stock Shortage Specialist
When a component is unavailable, discontinued or has an unacceptable lead time, we tap into our network of vetted European and Asian distributors to source what you need — without compromising on quality or traceability.
Request a quote →Compliant Alternatives
We identify pin-to-pin, electrically equivalent substitutes that meet the same certifications (RoHS, AEC-Q100, REACH) as your original specification — validated against datasheets, not just part numbers. Often at a lower cost.
BOM Analysis service →